THE CALIBRATION OF FLOW IONISATION CHAMBERS FOR TRITIUM MONITORING IN AIR
Technical Report
·
OSTI ID:4201473
A stream of air of knowu tritium (HTO) concentration is obtained by mixing dry air and air saturated with HTO vapor in knowu proportions. This was used to calibrate ion chambers for the monitoring of tritium in air. Particular attention is paid io the 1590A, a chamber designed at A.E.R.E. Harwell, and limitations in its use as an air monitor due to adsorption of HTO are briefly discussed. (auth)
- Research Organization:
- United Kingdom Atomic Energy Authority. Research Group. Atomic Energy Research Establishment, Harwell, Berks, England
- NSA Number:
- NSA-14-000304
- OSTI ID:
- 4201473
- Report Number(s):
- AERE-M-429
- Country of Publication:
- United Kingdom
- Language:
- English
Similar Records
Evaluation of the response of tritium-in-air instrumentation to HT in dry and humid conditions and to HTO vapor
Tritium monitor calibration at Los Alamos National Laboratory
Tritium measurement and monitoring in experimental and process systems with ionization chambers
Journal Article
·
Sun Mar 15 00:00:00 EDT 2015
· Fusion Science and Technology
·
OSTI ID:22429757
Tritium monitor calibration at Los Alamos National Laboratory
Technical Report
·
Fri Aug 01 00:00:00 EDT 1997
·
OSTI ID:634020
Tritium measurement and monitoring in experimental and process systems with ionization chambers
Conference
·
Sat Feb 29 23:00:00 EST 1992
· Fusion Technology; (United States)
·
OSTI ID:5087029