Evaluation of the response of tritium-in-air instrumentation to HT in dry and humid conditions and to HTO vapor
Journal Article
·
· Fusion Science and Technology
- National Physical Laboratory, Teddington (United Kingdom)
- Grenoble Institute of Technology - Phelma, Grenoble (France)
Nuclear plant operators (power generation, decommissioning and reprocessing operations) are required to monitor releases of tritium species for regulatory compliance and radiation protection purposes. Tritium monitoring is performed using tritium-in-air gas monitoring instrumentation based either on flow-through ion chambers or proportional counting systems. Tritium-in-air monitors are typically calibrated in dry conditions but in service may operate at elevated levels of relative humidity. The NPL (National Physical Laboratory) radioactive gas-in-air calibration system has been used to study the effect of humidity on the response to tritium of two tritium-in-air ion chamber based monitors and one proportional counting system which uses a P10/air gas mixture. The response of these instruments to HTO vapour has also been evaluated. In each case, instrument responses were obtained for HT in dry conditions (relative humidity (RH) about 2%), HT in 45% RH, and finally HTO at 45% RH. Instrumentation response to HT in humid conditions has been found to slightly exceed that in dry conditions. (authors)
- OSTI ID:
- 22429757
- Journal Information:
- Fusion Science and Technology, Journal Name: Fusion Science and Technology Journal Issue: 3 Vol. 67; ISSN 1536-1055
- Country of Publication:
- United States
- Language:
- English
Similar Records
Differential measurement of low level HTO and HT leak rates
An evaluation of the CRNL/CFFTP HT/HTO discriminating monitor
A Passive Diffusion Sampler for HT- and HTO-in-Air
Conference
·
Sun Aug 01 00:00:00 EDT 1976
·
OSTI ID:7322691
An evaluation of the CRNL/CFFTP HT/HTO discriminating monitor
Journal Article
·
Thu Sep 01 00:00:00 EDT 1988
· Fusion Technol.; (United States)
·
OSTI ID:6135298
A Passive Diffusion Sampler for HT- and HTO-in-Air
Journal Article
·
Fri Jul 15 00:00:00 EDT 2005
· Fusion Science and Technology
·
OSTI ID:20854173