ON THE METHOD OF ABSOLUTE MEASUREMENTS OF ACTIVITY IN GAS LABELED WITH C$sup 14$ AND S$sup 3$$sup 5$ (in Russian)
Journal Article
·
· Pribory i Tekh. Ekspt.
OSTI ID:4159972
A method is suggested for measuring the activity of C/sup 14/and S/sup 35/-labeled gas and vapors with a "free gas" ionization chamber. An analysis is made of the utilization of heavy inert gas (xenon or freon-12) under 1 to 3 atm pressure as the main filler with approximately 0.5% of the gas to be measured. Such a method demands small chamber dimensions (diameter and height approximately 100 mm at 3 atm pressure). The order of accuracy is approximately plus or minus 5%. A schematic drawing of the chamber is included. (tr-auth)
- Research Organization:
- Originating Research Org. not identified
- NSA Number:
- NSA-14-023064
- OSTI ID:
- 4159972
- Journal Information:
- Pribory i Tekh. Ekspt., Journal Name: Pribory i Tekh. Ekspt. Vol. Vol: No. 3
- Country of Publication:
- Country unknown/Code not available
- Language:
- Russian