SOLUTION TECHNIQUE IN ANALYSIS OF METAL ALLOYS BY X-RAY FLUORESCENT SPECTROMETRY
Technical Report
·
OSTI ID:4147605
In a study of the effect of variations in the amounts of diverse metal and acid ions present in a solution of metal alloy to be analyzed by x-ray fluorescence there was no appreciable metal-metal interference, other than direct line interference, in any practical case. There was, however, signal attenuation by all common solvent acids: absolute acid concentration must therefore be controlled to tenths of a milliliter for any quantitative work. (auth)
- Research Organization:
- Nuclear Metals, Inc., Concord, Mass.
- NSA Number:
- NSA-14-011901
- OSTI ID:
- 4147605
- Report Number(s):
- NMI-1231
- Country of Publication:
- United States
- Language:
- English
Similar Records
X-RAY FLUORESCENCE METHOD FOR DETERMINATION OF THE TRANSITION-METAL CONTENT IN VERY SMALL SPECIMENS OF ALLOYS
X-RAY FLUORESCENCE ANALYSIS OF PLUTONIUM
Information: Use of the Ex-6500 energy-dispersive x-ray fluorescence spectrometer in quantitative analysis. Component determination in noble metal alloys
Journal Article
·
Sat Oct 01 00:00:00 EDT 1960
· J. Sci. Instr.
·
OSTI ID:4143944
X-RAY FLUORESCENCE ANALYSIS OF PLUTONIUM
Journal Article
·
Thu Dec 31 23:00:00 EST 1959
· Talanta (England)
·
OSTI ID:4102833
Information: Use of the Ex-6500 energy-dispersive x-ray fluorescence spectrometer in quantitative analysis. Component determination in noble metal alloys
Journal Article
·
Tue Oct 31 23:00:00 EST 1995
· Industrial Laboratory
·
OSTI ID:244126