AN EXPERIMENTAL RADON-IN-AIR MONITOR

Technical Report ·
OSTI ID:4139674
An instrument was designed to detect radon in air at or about a level of 10/sup -13/ curies of radon per cm/sup 3/ of air. The 5.5 Mev alpha particle emitted in the decay of radon (/sub 86/Rn/sup 222/) to radium A (/sub 84/ Po/sup 218/) is detected by a large screen scintillation counter. The instrument measures the pulse rate, due to these alpha particles, from the scintillation counter and hence the radon-in-air concentration. The measured sensitivity for 10/sup -13/ curies of radon per cm/sup 3/ of air is 5 counts per second; the background level is less than 0.4 counts per second. The activity level is displayed on a meter in terms of counts per second on successive ranges from 1 to 10/sup 5/ counts per second full scale and facilities are available for operation of a chart recorder. An audible warning circuit is incorporated and this may be set to operate at any convenient level. (auth)
Research Organization:
United Kingdom Atomic Energy Authority. Research Group. Atomic Energy Research Establishment, Harwell, Berks, England
NSA Number:
NSA-14-019125
OSTI ID:
4139674
Report Number(s):
AERE-M-660
Country of Publication:
United Kingdom
Language:
English

Similar Records

A CALIBRATION TECHNIQUE FOR THE DETERMINATION OF RADON-222 AND ITS APPLICATION TO VARIOUS MEASURING INSTRUMENTS
Technical Report · Mon Dec 31 23:00:00 EST 1962 · OSTI ID:4747784

Radon counting statistics - a Monte Carlo investigation
Journal Article · Sat Jun 01 00:00:00 EDT 1996 · Health Physics · OSTI ID:394062

METHODS FOR DETERMINING RADON-222 AND RADIUM-226
Journal Article · Sat Dec 31 23:00:00 EST 1960 · Journal American Water Works Association (U.S.) · OSTI ID:4103582