AN ULTRA-HIGH VACUUM SYSTEM USING AN OIL-DIFFUSION PUMP WITH A NON- REFRIGERATED ISOLATION TRAP

Journal Article · · Experientia (Switzerland)
DOI:https://doi.org/10.1007/BF02171160· OSTI ID:4130284
Ultrahigh vacuums with a basic pressure of less than 5 x l0/sup -10/ mm Hg were routinely produced and maintained for several months by means of an oil diffusion pump. An isolation trap charged with activated alumina of large cross section is placed between pump and glass system. The back flow of oil vapors into the experimental apparatus was completely prevented without cooling. (tr- auth)
Research Organization:
Stanford Research Inst., Menlo Park, Calif.
Sponsoring Organization:
USDOE
NSA Number:
NSA-15-002724
OSTI ID:
4130284
Journal Information:
Experientia (Switzerland), Journal Name: Experientia (Switzerland) Vol. Vol: 16; ISSN EXPEA
Country of Publication:
Country unknown/Code not available
Language:
English

Similar Records

HIGH-SPEED NONREFRIGERATED ISOLATION TRAPS FOR ULTRA HIGH-VACUUM SYSTEMS
Journal Article · Tue Sep 01 00:00:00 EDT 1959 · Rev. Sci. Instr. · OSTI ID:4213738

ELECTRONIC ULTRA-HIGH VACUUM PUMP
Journal Article · Thu May 01 00:00:00 EDT 1958 · Rev. Sci. Instr. · OSTI ID:4303603

NONPROPORTIONALITY IN BAYARD-ALPERT IONIZATION GAUGE AND THE ULTIMATE VACUUM DETERMINATION OF DIFFUSION PUMPS