THE STAGE CONSTANTS OF CASCADE IMPACTORS. Technical Progress Report

Technical Report ·
OSTI ID:4128529

The particle size distributions which would be found on each stage of a csscade impactor when used to sample and aerosol having a lognormal mass distribution were calculated. Three impactors, differing with respect to the collection characteristics of the various stages, were assumed. Samples were obtained from hypothetical aerosols having a wide randge of distribution parameters. It is shown that for a particular aerosol the particle sizes appear to be lognormally distributed on all stages and that the mass mediand diameter for a given stage, calculated from the number distribution on that stsge, is a good estimate of the true mass median diameter for that stage. The true mass median diameters and true effective cut-off diameters for each stage and for each aerosol were calculated also. The latter are shown to be superior for interpreting impactor data. (auth)

Research Organization:
Lovelace Foundation for Medical Education and Research, Albuquerque, N.M.
NSA Number:
NSA-18-004052
OSTI ID:
4128529
Report Number(s):
LF-12
Country of Publication:
United States
Language:
English

Similar Records

Personal cascade impactor: design, evaluation and calibration
Journal Article · Mon Jun 01 00:00:00 EDT 1987 · Am. Ind. Hyg. Assoc. J.; (United States) · OSTI ID:5671323

Calibration and evaluation of a real-time cascade impactor
Journal Article · Sat Mar 31 23:00:00 EST 1984 · Am. Ind. Hyg. Assoc. J.; (United States) · OSTI ID:5864809

Sampling charged particles with cascade impactors. Final report, April 1977-September 1978
Technical Report · Sun Dec 31 23:00:00 EST 1978 · OSTI ID:5643645