Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Personal cascade impactor: design, evaluation and calibration

Journal Article · · Am. Ind. Hyg. Assoc. J.; (United States)
A cascade impactor has been developed that is suitable for personal sampling. The impactor can be used with four, six, or eight stages with cut points ranging from 0.5 ..mu..m to 21 ..mu..m aerodynamic diameter. Although the impactor can be operated over a fairly wide flow rate range, it is designed specifically to be operated at 2 L/min so that a personal sampler pump may be used as the air mover. The nozzles of the impactor are radial slots with six slots per stage for the first six stages. Circular nozzles arranged in a radial pattern are used for the last two stages. On each stage, the area between the nozzles is used as the impaction surface for the stage before it. This design allows for the impactor to be compact, rugged and lightweight. The impactor stages are contained in a mounting bracket that can be clipped to the lapel or pocket. Although the impactor is designed for personal use, it also can be used as a compact impactor for general sampling tasks. The particle cut-off characteristics of each stage and the interstage losses were determined using monodisperse aerosols. The particle losses were found to be low and the cut-off characteristics sharp. Thus, size distributions can be obtained as accurately as with larger cascade impactors.
Research Organization:
Univ. of Minnesota, Minneapolis
OSTI ID:
5671323
Journal Information:
Am. Ind. Hyg. Assoc. J.; (United States), Journal Name: Am. Ind. Hyg. Assoc. J.; (United States) Vol. 48:6; ISSN AIHAA
Country of Publication:
United States
Language:
English

Similar Records

Stage response of an Andersen cascade impactor to monodisperse droplets
Technical Report · Mon May 01 00:00:00 EDT 1989 · OSTI ID:5882791

Calibration of a modified Sierra Model 235 slotted cascade impactor
Technical Report · Sun Jul 01 00:00:00 EDT 1979 · OSTI ID:5798734

Low flow rate sharp cut impactors for indoor air sampling: design and calibration
Journal Article · Sat Oct 31 23:00:00 EST 1987 · J. Air Pollut. Control Assoc.; (United States) · OSTI ID:5319742