A SIMPLE PLASTIC SCINTILLATION COUNTER FOR TRITIATED HYDROGEN
Journal Article
·
· Intern. J. Appl. Radiation and Isotopes
A plastic scintillation counter was designed and constructed for measuring the H/sup 3/ content of hydrogen emerging from a palladium chromatographic column. The counter consists of a piece of plastic phosphor shaped into a tube and inserted in the glass tubing loading from the column, with painted capillary tubing ends to exclude stray light. The sensitive volume of the detector is 0.5 ml, and a counting efficiency of 2.2% with a background of 5 c/sec at 20 ts C was obtained for tritiated hydrogen at 66 cm Hg. (D.L.C.)
- Research Organization:
- Wantage Research Lab., Berks, Eng.
- NSA Number:
- NSA-15-009073
- OSTI ID:
- 4105158
- Journal Information:
- Intern. J. Appl. Radiation and Isotopes, Journal Name: Intern. J. Appl. Radiation and Isotopes Vol. Vol: 9
- Country of Publication:
- Country unknown/Code not available
- Language:
- English
Similar Records
AGEING OF PLASTIC SCINTILLATORS
DETERMINATION OF TRITIUM IN A PROPORTIONAL COUNTER
A SIMPLE LIQUID SCINTILLATION COUNTER FOR MEASUREMENT OF C$sup 14$-LABELED COMPOUNDS