STATISTICAL ASPECTS OF DOUBLE ISOTOPE LIQUID SCINTILLATION COUNTING BY INTERNAL STANDARD TECHNIQUE

Journal Article · · Analytical Chemistry (U.S.) Formerly Ind. Eng. Chem., Anal. Ed.
OSTI ID:4032198
Equations were derived for calculation of the error in the disintegration rates of two isotopes in a mixture. The way in which the equations can be used to determine useful internal standard activities, channel settings, and isotope ratios is shown. For mixtures of carbon-14 and tritium, internal standard activities of 50,000-- 100,000 dpm for carbon-14 and 100,000-- 150,000 dpm for tritium are useful as single routine quantities. A 1-minute counting time after internal standard addition is routinely adequate. (auth)
Research Organization:
Lilly Research Labs., Indianapolis
NSA Number:
NSA-18-019872
OSTI ID:
4032198
Journal Information:
Analytical Chemistry (U.S.) Formerly Ind. Eng. Chem., Anal. Ed., Journal Name: Analytical Chemistry (U.S.) Formerly Ind. Eng. Chem., Anal. Ed. Vol. Vol: 36; ISSN ANCHA
Country of Publication:
Country unknown/Code not available
Language:
English

Similar Records

Statistical Aspects of Liquid Scintillation Counting by Internal Standard Technique. Single Isotope.
Journal Article · Sat Jun 01 00:00:00 EDT 1963 · Analytical Chemistry · OSTI ID:4709314

Liquid scintillation counting of /sup 14/C for differentiation of synthetic ethanol from ethanol of fermentation
Journal Article · Tue Sep 01 00:00:00 EDT 1981 · J. Assoc. Off. Anal. Chem.; (United States) · OSTI ID:6240169

CONTINUOUS SCINTILLATION COUNTING OF CARBON-14 AND TRITIUM IN EFFLUENT OF THE AUTOMATIC AMINO ACID ANALYZER
Journal Article · Fri Nov 30 23:00:00 EST 1962 · Analytical Biochemistry (U.S.) · OSTI ID:4739772