Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Post-injection transmission scans in a PET camera operating without septa with simultaneous measurement of emission activity contamination

Journal Article · · IEEE Transactions on Nuclear Science
DOI:https://doi.org/10.1109/23.531883· OSTI ID:379758
;  [1]
  1. Univ. of Pennsylvania, Philadelphia, PA (United States). Dept. of Radiology

The authors report here on methods developed to reliably perform attenuation correction by post-injection transmission in a volume imaging PET scanner. The method directly measures the emission contamination during the transmission study, using a virtual transmission source position offset by 20{degree} from the actual {sup 68}Ge transmission source. Events are recorded only if they meet a co-linearity requirement with either the real or virtual source position. The simultaneous measurements of the emission contamination and transmission data remove the need for complex corrections to the emission data and are not subject to activity redistributions between emission and transmission scans performed sequentially. Correction is necessary, however, for the extra deadtime that varies with the amount of emission activity in the FOV. The extra deadtime is determined from a lookup table of deadtime as a function of detector countrates, which are recorded during the study. The lookup table is based upon phantom measurements. Using patient and phantom data, with both pre- and post-injection transmission measurements, this method is shown to be reliable for attenuation correction in the body. In addition, it enables the calculation of Standardized Uptake Values for analyzing tumor activity.

OSTI ID:
379758
Report Number(s):
CONF-951073--
Journal Information:
IEEE Transactions on Nuclear Science, Journal Name: IEEE Transactions on Nuclear Science Journal Issue: 4Pt1 Vol. 43; ISSN 0018-9499; ISSN IETNAE
Country of Publication:
United States
Language:
English

Similar Records

Post injection transmission scanning in a volume imaging PET camera
Conference · Mon Aug 01 00:00:00 EDT 1994 · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States) · OSTI ID:7124051

A practical method for randoms subtraction in volume imaging PET from detector singles countrate measurements
Journal Article · Sat Jun 01 00:00:00 EDT 1996 · IEEE Transactions on Nuclear Science · OSTI ID:276530

A comparison of segmentation and emission subtraction for singles transmission in PET
Journal Article · Mon Jun 01 00:00:00 EDT 1998 · IEEE Transactions on Nuclear Science · OSTI ID:649400