Title: Specimen preparation for x-ray fluorescence analysis of solutions

Journal Article · · Advances in X-Ray Analysis

Specimens for x-ray fluorescence analysis (XRFA) were prepared by adding dry gelatine (10%) to the analysis solution, homogenizing the mixture and cooling for 20 minutes. Thus, a compact resilient mass could be formed with the required shape and size; the roughness of the surface was determined by the roughness of the surface on which the specimen was formed, much the same as highly polished. Various calibration methods can be applied in the XRFA of a variety of materials if such specimens are used. 12 refs., 1 fig., 2 tabs.

Sponsoring Organization:
USDOE
OSTI ID:
369877
Report Number(s):
CONF-9408178-; ISSN 0376-0308; TRN: 96:004756-0081
Journal Information:
Advances in X-Ray Analysis, Vol. 38; Conference: 43. annual Denver x-ray conference on applications of x-ray analysis, Steamboat Spring, CO (United States), 1-5 Aug 1994; Other Information: PBD: 1995
Country of Publication:
United States
Language:
English

Similar Records

Transmission-corrected x-ray fluorescence analysis of uranium and plutonium solutions using a dual transmission source
Conference · Thu Jan 01 00:00:00 EST 1987 · Transactions of the American Nuclear Society; (USA) · OSTI ID:369877

A transportable x-ray fluorescence analysis system for assay of plutonium and uranium solutions in gloveboxes
Conference · Fri Jun 01 00:00:00 EDT 1990 · OSTI ID:369877

X-ray fluorescence analysis of ACHEX aerosols
Journal Article · Tue Jan 01 00:00:00 EST 1980 · Adv. Environ. Sci. Technol.; (United States) · OSTI ID:369877