Laser-induced breakdown spectrometry as a multimetal continuous-emission monitor
- Diagnostic Instrumentation and Analysis Laboratory, Mississippi State University, 205 Research Boulevard, Mississippi Research and Technology Park, Starkville, Mississippi 39759-9734 (United States)
Laser-induced breakdown spectrometry (LIBS) has been used to detect atomic and molecular species in various environments. LIBS has the capability to be used as a continuous-emission monitor to monitor toxic-metal concentrations in stack emissions. Recently a mobile LIBS system was calibrated in our laboratory and tested as a multimetal continuous-emission monitor during a joint U.S. Department of Energy{endash}Environmental Protection Agency (EPA) test. LIBS measurements were performed with three sets of metal concentrations at the EPA Rotary Kiln Incinerator Simulator. The LIBS system successfully measured concentrations of Cr, Pb, Cd, and Be in near real time in this test. Real-time LIBS data were averaged and compared with data obtained from an EPA reference method that was conducted concurrently with LIBS. The details of the LIBS calibration and results of these LIBS measurements are described. {copyright} 1999 Optical Society of America
- DOE Contract Number:
- FG21-93CH10575
- OSTI ID:
- 365850
- Journal Information:
- Applied Optics, Journal Name: Applied Optics Journal Issue: 9 Vol. 38; ISSN APOPAI; ISSN 0003-6935
- Country of Publication:
- United States
- Language:
- English
Similar Records
Performance testing of multi-metal continuous emissions monitors
Continuous emissions monitoring demonstration program