Electrical discharge machining of type-N(f) microwave connectors

Technical Report ·
DOI:https://doi.org/10.2172/285503· OSTI ID:285503

A particular out-of-specification mechanical dimension on Type-N(f) [Type-N(female)] microwave connectors sometimes disqualifies otherwise perfectly acceptable microwave devices from being used in calibration systems. The Miniature Machining Group at Sandia National Laboratories applied a technique called Electrical Discharge Machining (EDM) to quickly and economically machine these devices without disassembly. In so doing, they facilitated the use of existing components without the need to purchase new devices. The technique also improves an uncertainty of calibration known as Mismatch Uncertainty by optimizing the reflection coefficient of the calibration test port. This effects a reduction in overall calibration uncertainties.

Research Organization:
Sandia National Labs., Albuquerque, NM (United States)
Sponsoring Organization:
USDOE, Washington, DC (United States)
DOE Contract Number:
AC04-94AL85000
OSTI ID:
285503
Report Number(s):
SAND--96-1891; ON: DE96013853
Country of Publication:
United States
Language:
English

Similar Records

Cryogenic single-port calibration for superconducting microwave resonator measurements
Journal Article · Mon Jun 28 00:00:00 EDT 2021 · Quantum Science and Technology · OSTI ID:1810671

A calibration method of microwave noise sources with different types of connectors
Conference · Fri Dec 30 23:00:00 EST 1994 · OSTI ID:125772

Printed Microwave Connector
Conference · Wed Jun 01 00:00:00 EDT 2022 · OSTI ID:1846012