Scintillation and cherenkov photon counting detectors with analog silicon photomultipliers for TOF-PET

Journal Article · · Physics in Medicine and Biology
 [1];  [1];  [2]
  1. Lawrence Berkeley National Laboratory (LBNL), Berkeley, CA (United States)
  2. Sandia National Laboratories (SNL-CA), Livermore, CA (United States)
Objective. Standard signal processing approaches for scintillation detectors in positron emission tomography (PET) derive accurate estimates for 511 keV photon time of interaction and energy imparted to the detection media from aggregate characteristics of electronic pulse shapes. The ultimate realization of a scintillation detector for PET is one that provides a unique timestamp and position for each detected scintillation photon. Detectors with these capabilities enable advanced concepts for three-dimensional (3D) position and time of interaction estimation with methods that exploit the spatiotemporal arrival time kinetics of individual scintillation photons. Approach. In this work, we show that taking into consideration the temporal photon emission density of a scintillator, the channel density of an analog silicon photomultiplier (SiPM) array, and employing fast electronic readout with digital signal processing, a detector that counts and timestamps scintillation photons can be realized. To demonstrate this approach, a prototype detector was constructed, comprising multichannel electronic readout for a bismuth germanate (BGO) scintillator coupled to an SiPM array. Main Results. In proof-of-concept measurements with this detector, we were able to count and provide unique timestamps for 66% of all optical photons, where the remaining 34% (two-or-more-photon pulses) are also independently counted, but each photon bunch shares a common timestamp. We show this detector concept can implement 3D positioning of 511 keV photon interactions and thereby enable corrections for time of interaction estimators. The detector achieved 17.6% energy resolution at 511 keV and 237 ± 10 ps full-width-at-half-maximum coincidence time resolution (CTR) (fast spectral component) versus a reference detector. We outline the methodology, readout, and approach for achieving this detector capability in first-ever, proof-of-concept measurements for scintillation photon counting detector with analog silicon photomultipliers. Significance. The presented detector concept is a promising design for large area, high sensitivity TOF-PET detector modules that can implement advanced event positioning and time of interaction estimators, which could push state-of-the-art performance.
Research Organization:
Sandia National Laboratories (SNL-CA), Livermore, CA (United States)
Sponsoring Organization:
National Institutes of Health (NIH); USDOE; USDOE National Nuclear Security Administration (NNSA)
Grant/Contract Number:
NA0003525
OSTI ID:
2471747
Alternate ID(s):
OSTI ID: 2584906
Report Number(s):
SAND--2025-07683J
Journal Information:
Physics in Medicine and Biology, Journal Name: Physics in Medicine and Biology Journal Issue: 4 Vol. 69; ISSN 0031-9155
Publisher:
IOP PublishingCopyright Statement
Country of Publication:
United States
Language:
English

References (33)

Effect of SSPM surface coating on light collection efficiency and optical crosstalk for scintillation detection
  • Barton, Paul; Stapels, Christopher; Johnson, Erik
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 610, Issue 1 https://doi.org/10.1016/j.nima.2009.05.187
journal October 2009
Time resolution deterioration with increasing crystal length in a TOF-PET system
  • Gundacker, S.; Knapitsch, A.; Auffray, E.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 737 https://doi.org/10.1016/j.nima.2013.11.025
journal February 2014
A new time calibration method for switched-capacitor-array-based waveform samplers
  • Kim, H.; Chen, C. -T.; Eclov, N.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 767 https://doi.org/10.1016/j.nima.2014.08.025
journal December 2014
Advances in digital SiPMs and their application in biomedical imaging
  • Schaart, Dennis R.; Charbon, Edoardo; Frach, Thomas
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 809 https://doi.org/10.1016/j.nima.2015.10.078
journal February 2016
Precise rise and decay time measurements of inorganic scintillators by means of X-ray and 511 keV excitation
  • Gundacker, S.; Turtos, R. M.; Auffray, E.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 891 https://doi.org/10.1016/j.nima.2018.02.074
journal May 2018
Fully Digital Arrays of Silicon Photomultipliers (dSiPM) – a Scalable Alternative to Vacuum Photomultiplier Tubes (PMT) journal January 2012
Sub-200 ps CRT in monolithic scintillator PET detectors using digital SiPM arrays and maximum likelihood interaction time estimation journal April 2013
SiPM optical crosstalk amplification due to scintillator crystal: effects on timing performance journal June 2014
Time-based position estimation in monolithic scintillator detectors journal July 2015
Towards monolithic scintillator based TOF-PET systems: practical methods for detector calibration and operation journal June 2016
Bismuth germanate coupled to near ultraviolet silicon photomultipliers for time-of-flight PET journal September 2016
BGO as a hybrid scintillator / Cherenkov radiator for cost-effective time-of-flight PET journal May 2017
Sub-3 mm, near-200 ps TOF/DOI-PET imaging with monolithic scintillator detectors in a 70 cm diameter tomographic setup journal July 2018
Improved single photon time resolution for analog SiPMs with front end readout that reduces influence of electronic noise journal September 2018
High-frequency SiPM readout advances measured coincidence time resolution limits in TOF-PET journal February 2019
Electronics method to advance the coincidence time resolution with bismuth germanate journal September 2019
Experimental time resolution limits of modern SiPMs and TOF-PET detectors exploring different scintillators and Cherenkov emission journal January 2020
Pushing Cherenkov PET with BGO via coincidence time resolution classification and correction journal June 2020
Roadmap toward the 10 ps time-of-flight PET challenge journal October 2020
DOI estimation through signal arrival time distribution: a theoretical description including proof of concept measurements journal April 2021
On timing-optimized SiPMs for Cherenkov detection to boost low cost time-of-flight PET journal August 2023
Position algorithm for monolithic scintillation crystals based on charge projection readout journal January 2016
Single photon time resolution of state of the art SiPMs journal October 2016
A 4 × 4 × 416 digital SiPM array with 192 TDCs for multiple high-resolution timestamp acquisition journal May 2013
Measurement of LYSO Intrinsic Light Yield Using Electron Excitation journal April 2016
Exploring Cherenkov Emission of BGO for TOF-PET journal September 2021
Evolution of PET Detectors and Event Positioning Algorithms Using Monolithic Scintillation Crystals journal May 2021
Timing Estimation and Limits in TOF-PET Detectors Producing Prompt Photons journal September 2023
Suppression of the optical crosstalk in a multi-channel silicon photomultiplier array journal January 2021
Studies of a Next-Generation Silicon-Photomultiplier–Based Time-of-Flight PET/CT System journal April 2017
Performance Characteristics of the Digital Biograph Vision PET/CT System journal January 2019
Cherenkov Radiation–Based Coincidence Time Resolution Measurements in BGO Scintillators journal January 2022
3D Photon-To-Digital Converter for Radiation Instrumentation: Motivation and Future Works journal January 2021