Front-End Design for SiPM-Based Monolithic Neutron Double Scatter Imagers

Journal Article · · Sensors
DOI:https://doi.org/10.3390/s22093553· OSTI ID:1877164
 [1];  [2];  [2];  [1];  [3];  [3];  [2]
  1. Lawrence Berkeley National Lab. (LBNL), Berkeley, CA (United States)
  2. Sandia National Lab. (SNL-CA), Livermore, CA (United States)
  3. Oak Ridge National Lab. (ORNL), Oak Ridge, TN (United States)

Neutron double scatter imaging exploits the kinematics of neutron elastic scattering to enable emission imaging of neutron sources. Due to the relatively low coincidence detection efficiency of fast neutrons in organic scintillator arrays, imaging efficiency for double scatter cameras can also be low. One method to realize significant gains in neutron coincidence detection efficiency is to develop neutron double scatter detectors which employ monolithic blocks of organic scintillator, instrumented with photosensor arrays on multiple faces to enable 3D position and multi-interaction time pickoff. Silicon photomultipliers (SiPMs) have several advantageous characteristics for this approach, including high photon detection efficiency (PDE), good single photon time resolution (SPTR), high gain that translates to single photon counting capabilities, and ability to be tiled into large arrays with high packing fraction and photosensitive area fill factor. However, they also have a tradeoff in high uncorrelated and correlated noise rates (dark counts from thermionic emissions and optical photon crosstalk generated during avalanche) which may complicate event positioning algorithms. We have evaluated the noise characteristics and SPTR of Hamamatsu S13360-6075 SiPMs with low noise, fast electronic readout for integration into a monolithic neutron scatter camera prototype. The sensors and electronic readout were implemented in a small-scale prototype detector in order to estimate expected noise performance for a monolithic neutron scatter camera and perform proof-of-concept measurements for scintillation photon counting and three-dimensional event positioning.

Research Organization:
Sandia National Laboratories (SNL-NM), Albuquerque, NM (United States); Sandia National Laboratories (SNL-CA), Livermore, CA (United States)
Sponsoring Organization:
USDOE National Nuclear Security Administration (NNSA)
Grant/Contract Number:
NA0003525
OSTI ID:
1877164
Report Number(s):
SAND2022-9374J; 708149
Journal Information:
Sensors, Journal Name: Sensors Journal Issue: 9 Vol. 22; ISSN 1424-8220
Publisher:
MDPI AGCopyright Statement
Country of Publication:
United States
Language:
English

References (13)

Effect of SSPM surface coating on light collection efficiency and optical crosstalk for scintillation detection
  • Barton, Paul; Stapels, Christopher; Johnson, Erik
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 610, Issue 1 https://doi.org/10.1016/j.nima.2009.05.187
journal October 2009
Analysis of single-photon time resolution of FBK silicon photomultipliers
  • Acerbi, Fabio; Ferri, Alessandro; Gola, Alberto
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 787 https://doi.org/10.1016/j.nima.2014.10.057
journal July 2015
Characterization of three high efficiency and blue sensitive silicon photomultipliers
  • Otte, Adam Nepomuk; Garcia, Distefano; Nguyen, Thanh
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 846 https://doi.org/10.1016/j.nima.2016.09.053
journal February 2017
Improved single photon time resolution for analog SiPMs with front end readout that reduces influence of electronic noise journal September 2018
High-frequency SiPM readout advances measured coincidence time resolution limits in TOF-PET journal February 2019
Single photon time resolution of state of the art SiPMs journal October 2016
SCEMA: a high channel density electronics module for fast waveform capture journal February 2019
Results With the Neutron Scatter Camera journal June 2009
A Comprehensive Model of the Response of Silicon Photomultipliers journal August 2010
Dark Count Impact for First Photon Discriminators for SPAD Digital Arrays in PET journal June 2015
Dark Count Resilient Time Estimators for Time-of-Flight PET journal January 2020
Evolution of PET Detectors and Event Positioning Algorithms Using Monolithic Scintillation Crystals journal May 2021
Suppression of the optical crosstalk in a multi-channel silicon photomultiplier array journal January 2021

Similar Records

Characterization of a SiPM-based monolithic neutron scatter camera using dark counts
Journal Article · Tue Jun 11 00:00:00 EDT 2024 · Journal of Instrumentation · OSTI ID:2426036

Optimizing the spatial resolution and gamma discrimination of SiPM-based Anger cameras
Journal Article · Mon Sep 02 00:00:00 EDT 2024 · Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment · OSTI ID:2447302