Characteristics of the 2 m long scintillation counters measured by detecting 90 MeV protons

Journal Article · · Physics of Particles and Nuclei
; ;  [1]
  1. Institute for High Energy Physics, National Research Center Kurchatov Institute, Protvino, Moscow oblast (Russian Federation)
The characteristics of scintillation counters measured in a 90-MeV proton beam have been presented. Two-meter long scintillators BC-404 and BC-408, which were viewed from both ends with photoelectron multipliers R1828-01 (PMT) have been used. Both the amplitude and time distributions of the PMT signals are well described by the Gaussian function. It turned out that the amplitude resolution σ(А)/А weakly depends on the type of scintillator and the proton position along the counter, and lies in the interval between 4 and 5%. The best time resolution σ(T) is about 50 ps near the counter ends and 63 ps at the center.
OSTI ID:
22979601
Journal Information:
Physics of Particles and Nuclei, Journal Name: Physics of Particles and Nuclei Journal Issue: 1 Vol. 49; ISSN 1063-7796
Country of Publication:
United States
Language:
English

Similar Records

Studying the characteristics of scintillation counters in a beam of relativistic particles
Journal Article · Sun Jan 14 23:00:00 EST 2018 · Physics of Particles and Nuclei · OSTI ID:22979593

Scintillation counters characteristics versus light collection conditions
Journal Article · Sat Dec 31 23:00:00 EST 1994 · Instruments and Experimental Techniques · OSTI ID:123404

SCINTILLATION COUNTER
Journal Article · Fri Jan 31 23:00:00 EST 1964 · Kerntechnik (West Germany) Merged with Atomkernenergie to form Atomkernenerg./Kerntech.kyu Hokoku · OSTI ID:4053147