The effect of the light collection conditions on a scintillation counter`s performance have been studied by measuring the characteristics of a counter with a rectangular plastic scintillator surrounded by air, water, and paraffin oil, and those of a counter with trapezium scintillators with angles between side faces of 5.6, 7.5, and 10{degrees}. The measurements demonstrate that the time resolution is mainly defined by a number of collected photons. When the refraction index of the environment rises, the effective light propagation velocity towards the PMT rises. It is possible to select an angle between the faces of a trapezium scintillator at which the detected signal amplitude is virtually independent of the scintillation site.
OSTI ID:
123404
Journal Information:
Instruments and Experimental Techniques, Journal Name: Instruments and Experimental Techniques Journal Issue: 4 Vol. 37; ISSN INETAK; ISSN 0020-4412
Country of Publication:
United States
Language:
English

Similar Records

Increasing the Light Collection Efficiency of Scintillation Counters
Journal Article · Sat Jun 01 00:00:00 EDT 1963 · Review of Scientific Instruments · OSTI ID:4687598

INCREASING THE LIGHT COLLECTION EFFICIENCY OF SCINTILLATION COUNTERS
Technical Report · Mon Oct 01 00:00:00 EDT 1962 · OSTI ID:4731801

Light collection from scintillation counters using WLS fibers and bars
Journal Article · Sun Nov 08 23:00:00 EST 1998 · AIP Conference Proceedings · OSTI ID:21199299