Radiation Exposure During Transarterial Chemoembolization: Angio-CT Versus Cone-Beam CT

Journal Article · · Cardiovascular and Interventional Radiology
 [1];  [2]; ; ; ;  [1]; ;  [3];  [4];  [1]
  1. St-Eloi Hospital, CHU Montpellier, University of Montpellier, Department of Radiology (France)
  2. CHU Montpellier, Department of Medical Physics (France)
  3. CHU Nimes, University of Montpellier, Department of Radiology (France)
  4. IMAG, CNRS, University of Montpellier, CHU Montpellier (France)

Introduction: Cone-beam computed tomography (CBCT) has been developed to improve reliability of many interventional radiology (IR) procedures performed with Angio system, such as transarterial chemoembolization (TACE). Angio-CT has emerged as a new imaging technology that combines a CT scanner with an Angio system in the same IR suite. The purpose of our study was to compare Angio system with CBCT capability and Angio-CT in terms of patient radiation exposure during TACE procedures. Materials and Methods: Consecutive TACE procedures performed between January 2016 and September 2017 with the two imaging modalities (Artis Zeego defining the CBCT group and Infinix-i 4D-CT defining the Angio-CT group) were reviewed. TACE and patient’s characteristics and patient radiation exposure parameters were collected. Dose-area products (DAP) and dose-length products (DLP) were converted into effective doses (ED) using conversion factors. Accuracy of tumor targeting and response was retrospectively assessed. Results: A total of 114 TACE procedures in 96 patients were included with 57 procedures in each group. The total ED in the Angio-CT group was 2.5 times lower than that in the CBCT group (median 15.4 vs. 39.2 mSv, p < 0.001). Both 2D ED and 3D ED were lower in the Angio-CT group than in the CBCT group (5.1 vs. 20 mSv, p < 0.001, and 7.4 vs. 17.9 mSv, p < 0.001, respectively). There was no significant difference neither in terms of classes of tumor targeting (p = 0.509) nor in terms of classes of tumor response (p = 0.070) between both groups. Conclusion: Angio-CT provides significant decrease in patient effective dose during TACE procedures compared to Angio system with CBCT.

OSTI ID:
22970458
Journal Information:
Cardiovascular and Interventional Radiology, Journal Name: Cardiovascular and Interventional Radiology Journal Issue: 11 Vol. 42; ISSN 0174-1551; ISSN CAIRDG
Country of Publication:
United States
Language:
English

Similar Records

Percutaneous Bone Biopsies: Comparison between Flat-Panel Cone-Beam CT and CT-Scan Guidance
Journal Article · Sat Feb 14 23:00:00 EST 2015 · Cardiovascular and Interventional Radiology · OSTI ID:22470114

MDCT Versus MRI Assessment of Tumor Response After Transarterial Chemoembolization for the Treatment of Hepatocellular Carcinoma
Journal Article · Tue Jun 15 00:00:00 EDT 2010 · Cardiovascular and Interventional Radiology · OSTI ID:21428992

Efficacy and Safety of Transarterial Radioembolization Versus Chemoembolization in Patients With Hepatocellular Carcinoma
Journal Article · Sat Jun 15 00:00:00 EDT 2013 · Cardiovascular and Interventional Radiology · OSTI ID:22113655