Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Implementation of Precursor and Neutral Loss Scans on a Miniature Ion Trap Mass Spectrometer and Performance Comparison to a Benchtop Linear Ion Trap

Journal Article · · Journal of the American Society for Mass Spectrometry
Implementation of orthogonal double resonance precursor and neutral loss scans on the Mini 12 miniature rectilinear ion trap mass spectrometer is described, and performance is compared to that of a commercial Thermo linear trap quadropole (LTQ) linear ion trap. The ac frequency scan version of the technique at constant rf voltage is used here because it is operationally much simpler to implement. Remarkably, the Mini 12 shows up to two orders of magnitude higher sensitivity compared to that of the LTQ. Resolution on the LTQ is better than unit at scan speeds of ~ 400 Th/s, whereas peak widths on the Mini 12, on average, range from 0.5 to 2.0 Th full width at half maximum and depend heavily on the precursor ion Mathieu q parameter as well as the pump down time that precedes the mass scan. Both sensitivity and resolution are maximized under higher pressure conditions (short pump down time) on the Mini 12. The effective mass range of the product ion ejection waveform was found to be 5.8 Th on the Mini 12 in the precursor ion scan mode vs. that of 3.9 Th on the LTQ. In the neutral loss scan mode, the product ion selectivity was between 8 and 11 Th on the Mini 12 and between 7 and 8 Th on the LTQ. The effects of nonlinear resonance lines on the Mini 12 were also explored. .
OSTI ID:
22776895
Journal Information:
Journal of the American Society for Mass Spectrometry, Journal Name: Journal of the American Society for Mass Spectrometry Journal Issue: 7 Vol. 29; ISSN 1044-0305; ISSN JAMSEF
Country of Publication:
United States
Language:
English

Similar Records

Precursor and Neutral Loss Scans in an RF Scanning Linear Quadrupole Ion Trap
Journal Article · Sun Jul 15 00:00:00 EDT 2018 · Journal of the American Society for Mass Spectrometry · OSTI ID:22776896

Single Analyzer Precursor Ion Scans in a Linear Quadrupole Ion Trap Using Orthogonal Double Resonance Excitation
Journal Article · Fri Sep 15 00:00:00 EDT 2017 · Journal of the American Society for Mass Spectrometry · OSTI ID:22776906

Theory of high-resolution mass spectrometry achieved via resonance ejection in the quadrupole ion trap