SU-G-TeP1-03: Beam Quality Correction Factors for Linear Accelerator with and Without Flattening Filter

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4956993· OSTI ID:22649343
; ;  [1];  [2]
  1. Technische Hochschule Mittelhessen - University of Applied Sciences, Giessen (Germany)
  2. University Hospital for Medical Radiation Physics, Pius- Hospital, Medical Campus, Carl von Ossietzky University of Oldenburg (Germany)
Purpose: The impact of removing the flattening filter on absolute dosimetry based on IAEA’s TPR-398 and AAPM’s TG-51 was investigated in this study using Monte Carlo simulations. Methods: The EGSnrc software package was used for all Monte Carlo simulations performed in this work. Five different ionization chambers and nine linear accelerator heads have been modeled according to technical drawings. To generate a flattening filter free radiation field the flattening filter was replaced by a 2 mm thick aluminum layer. Dose calculation in a water phantom were performed to calculate the beam quality correction factor k{sub Q} as a function of the beam quality specifiers %dd(10){sub x}, TPR{sub 20,10} and mean photon and electron energies at the point of measurement in photon fields with (WFF) and without flattening filter (FFF). Results: The beam quality correction factor as a function of %dd(10){sub x} differs systematically between FFF and WFF beams for all investigated ionization chambers. The largest difference of 1.8% was observed for the largest investigated Farmer-type ionization chamber with a sensitive volume of 0.69 cm{sup 3}. For ionization chambers with a smaller nominal sensitive volume (0.015 – 0.3 cm{sup 3}) the deviation was less than 0.4% between WFF and FFF beams for %dd(10){sub x} > 62%. The specifier TPR{sub 20,10} revealed only a good correlation between WFF and FFF beams (< 0.3%) for low energies. Conclusion: The results confirm that %dd(10){sub x} is a suitable beam quality specifier for FFF beams with an acceptable bias. The deviation depends on the volume of the ionization chamber. Using %dd(10){sub x} to predict k{sub Q} for a large volume chamber in a FFF photon field may lead to not acceptable errors according to the results of this study. This bias may be caused by the volume effect due to the inhomogeneous photon fields of FFF linear accelerators.
OSTI ID:
22649343
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 6 Vol. 43; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

TH-E-BRE-06: Challenges in the Dosimetry of Flattening Filter Free Beams
Journal Article · Sun Jun 15 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22412396

Combining tissue-phantom ratios to provide a beam-quality specifier for flattening filter free photon beams
Journal Article · Sat Nov 01 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22320343

SU-E-T-525: Ionization Chamber Perturbation in Flattening Filter Free Beams
Journal Article · Mon Jun 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:22548563