Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

TH-E-BRE-06: Challenges in the Dosimetry of Flattening Filter Free Beams

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4889658· OSTI ID:22412396
; ;  [1]
  1. Technische Hochschule Mittelhessen - University of Applied Sciences, Giessen, Hesse (Germany)
Purpose: In current dosimetry protocols [AAPM TG51, IAEA TRS-389] the beam quality correction factor kQ and the water-to-air restricted mass collision stopping-power ratio SPR are related to beam quality specifiers %dd(10){sub x} respectively TPR{sub 20,10} Determining kQ and SPR using these regular beam quality specifiers for conventional accelerators (WFF) and flattening filter free accelerators (FFF) similarly could lead to systemic bias.The influence of the flattening filter on the relationship between various beam quality specifiers and SPR respectively k{sub Q} was studied using Monte Carlo simulations with realistic beam sources. Methods: All Monte Carlo simulations were performed using the BEAMnrc/EGSnrc code system. Radiation transport through nine linear accelerator heads modeled according to technical drawings given by the manufactures and a {sup 60} Co therapy source was simulated with BEAMnrc and then used as a radiation source for further simulations. FFF beam sources were implemented by removing the fattening filter from the WFF model. SPR was calculated applying the user code SPRRZnrc. The mean photon energy below the accelerator head and the mean energies of photons and electrons at the measuring point within the water phantom were calculated using FLURZnrc. Dose calculations within a small water voxel and the thimble ionization chamber PTW-31010 in a water depth of 10 cm were made using the egs-chamber code. Results: SPR and k{sub Q} as a function of fluence spectra based beam quality specifiers as well as conventional beam quality specifiers differ systematically between FFF and WFF beams. According to the results the specifier %dd(10){sub x} revealed the smallest deviation (max. 0.4%) between FFF and WFF beams. Conclusion: The results show that %dd(10){sub x} is an acceptable beam quality specifier for FFF beams. Nevertheless the results confirm the expected bias between FFF and WFF beams which must by further investigated.
OSTI ID:
22412396
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 6 Vol. 41; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

SU-G-TeP1-03: Beam Quality Correction Factors for Linear Accelerator with and Without Flattening Filter
Journal Article · Wed Jun 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22649343

Combining tissue-phantom ratios to provide a beam-quality specifier for flattening filter free photon beams
Journal Article · Sat Nov 01 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22320343

SU-F-T-488: Comparison of the TG-51 and TG-51 Addendum Calibration Protocols
Journal Article · Wed Jun 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22649075