SU-F-T-522: Dosimetric Study of Junction Dose in Double Isocenter Flatten and Flatten Filter Free IMRT and VMAT Plan Delivery
- Rajiv Gandhi Cancer Institute and Research Centre, New Delhi, Delhi (India)
Purpose: To quantify the dosimetric accuracy of junction dose in double isocenter flattened and flatten filter free(FFF) intensity modulated radiation therapy(IMRT) and volumetric modulated arc therapy(VMAT) plan delivery using pelvis phantom. Methods: Five large field pelvis patients were selected for this study. Double isocenter IMRT and VMAT treatment plans were generated in Eclipse Treatment planning System (V.11.0) using 6MV FB and FFF beams. For all the plans same distance 17.0cm was kept between one isocenter to another isocenter. IMRT Plans were made with 7 coplanar fields and VMAT plans were made with full double arcs. Dose calculation was performed using AAA algorithms with dose grid size of 0.25 cm. Verification plans were calculated on Scanditronix Wellhofer pelvis slab phantom. Measurement point was selected and calculated, where two isocenter plan fields are overlapping, this measurement point was kept at distance 8.5cm from both isocenter. The plans were delivered using Varian TrueBeamTM machine on pelvis slab phantom. Point dose measurements was carried out using CC13 ion chamber volume of 0.13cm3. Results: The measured junction point dose are compared with TPS calculated dose. The mean difference observed was 4.5%, 6.0%, 4.0% and 7.0% for IMRT-FB,IMRT-FFF, VMAT-FB and VMAT-FFF respectively. The measured dose results shows closer agreement with calculated dose in Flatten beam planning in both IMRT and VMAT, whereas in FFF beam plan dose difference are more compared with flatten beam plan. Conclusion: Dosimetry accuracy of Large Field junction dose difference was found less in Flatten beam compared with FFF beam plan delivery. Even though more dosimetric studies are required to analyse junction dose for FFF beam planning using multiple point dose measurements and fluence map verification in field junction area.
- OSTI ID:
- 22649108
- Journal Information:
- Medical Physics, Journal Name: Medical Physics Journal Issue: 6 Vol. 43; ISSN 0094-2405; ISSN MPHYA6
- Country of Publication:
- United States
- Language:
- English
Similar Records
Flattening Filter Free vs Flattened Beams for Breast Irradiation
Validation of a virtual source model for Monte Carlo dose calculations of a flattening filter free linac
SU-E-T-216: Comparison of Volumetrically Modulated Arc Therapy Treatment Using Flattening Filter Free Beams Vs. Flattened Beams for Partial Brain Irradiation
Journal Article
·
Thu Jan 31 23:00:00 EST 2013
· International Journal of Radiation Oncology, Biology and Physics
·
OSTI ID:22149782
Validation of a virtual source model for Monte Carlo dose calculations of a flattening filter free linac
Journal Article
·
Fri Jun 15 00:00:00 EDT 2012
· Medical Physics
·
OSTI ID:22098876
SU-E-T-216: Comparison of Volumetrically Modulated Arc Therapy Treatment Using Flattening Filter Free Beams Vs. Flattened Beams for Partial Brain Irradiation
Journal Article
·
Mon Jun 15 00:00:00 EDT 2015
· Medical Physics
·
OSTI ID:22548276