Flattening Filter Free vs Flattened Beams for Breast Irradiation
Journal Article
·
· International Journal of Radiation Oncology, Biology and Physics
Purpose: Flattening filter free (FFF) beams offer the potential for a higher dose rate, shorter treatment time, and lower peripheral dose. To investigate their role in large-field treatments, this study compared flattened and FFF beams for breast irradiation. Methods and Materials: Ten left breast clinical plans comprising 2 tangential beams and a medially located 3-field simultaneous integrated boost (SIB) were replanned. Full intensity modulated radiotherapy (IMRT), hybrid IMRT, electronic tissue compensator (ETC), and multiple static field treatment plans were created for the elective breast volume using flattened and FFF beams, in combination with a 3-field IMRT SIB. Plan quality was assessed and delivery times were measured for all plans for 1 patient. Out-of-field doses were measured using an ionization chamber for an IMRT plan optimized on a corner of simple cubic phantom for both flattened and FFF beams. Results: For each technique, mean target volume metrics (planning target volume coverage, homogeneity, conformity) were typically within 3% for flattened and FFF beams. Larger mean differences in boost conformity favoring flattened hybrid (7.2%) and full IMRT (5.5%) plans may have reflected limitations in plan normalization. Calculated heart and ipsilateral lung doses were comparable; however, both flattened and FFF low-dose phantom measurements were substantially higher than calculated values, rendering the comparison of low dose in the contralateral breast uncertain. Beam delivery times were on average 31% less for FFF. Conclusions: In general, target volume metrics for flattened and FFF plans were comparable. The planning system did not seem to allow for accurate peripheral dose evaluation. FFF was associated with a potentially shorter treatment time. All 4 IMRT techniques allowed FFF beams to generate acceptable plans for breast IMRT.
- OSTI ID:
- 22149782
- Journal Information:
- International Journal of Radiation Oncology, Biology and Physics, Journal Name: International Journal of Radiation Oncology, Biology and Physics Journal Issue: 2 Vol. 85; ISSN IOBPD3; ISSN 0360-3016
- Country of Publication:
- United States
- Language:
- English
Similar Records
SU-F-T-498: A Comparative Evaluation of 6MV Flatten Beam and Flattening Filter Free Photon Beam in Carcinoma Breast
SU-F-T-522: Dosimetric Study of Junction Dose in Double Isocenter Flatten and Flatten Filter Free IMRT and VMAT Plan Delivery
SU-F-T-372: Surface and Peripheral Dose in Compensator-Based FFF Beam IMRT
Journal Article
·
Wed Jun 15 00:00:00 EDT 2016
· Medical Physics
·
OSTI ID:22649085
SU-F-T-522: Dosimetric Study of Junction Dose in Double Isocenter Flatten and Flatten Filter Free IMRT and VMAT Plan Delivery
Journal Article
·
Wed Jun 15 00:00:00 EDT 2016
· Medical Physics
·
OSTI ID:22649108
SU-F-T-372: Surface and Peripheral Dose in Compensator-Based FFF Beam IMRT
Journal Article
·
Wed Jun 15 00:00:00 EDT 2016
· Medical Physics
·
OSTI ID:22648970