Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

First set of gated x-ray imaging diagnostics for the Laser Megajoule facility

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.4942930· OSTI ID:22597086

The Laser Megajoule (LMJ) facility located at CEA/CESTA started to operate in the early 2014 with two quadruplets (20 kJ at 351 nm) focused on target for the first experimental campaign. We present here the first set of gated x-ray imaging (GXI) diagnostics implemented on LMJ since mid-2014. This set consists of two imaging diagnostics with spatial, temporal, and broadband spectral resolution. These diagnostics will give basic measurements, during the entire life of the facility, such as position, structure, and balance of beams, but they will also be used to characterize gas filled target implosion symmetry and timing, to study x-ray radiography and hydrodynamic instabilities. The design requires a vulnerability approach, because components will operate in a harsh environment induced by neutron fluxes, gamma rays, debris, and shrapnel. Grazing incidence x-ray microscopes are fielded as far as possible away from the target to minimize potential damage and signal noise due to these sources. These imaging diagnostics incorporate microscopes with large source-to-optic distance and large size gated microchannel plate detectors. Microscopes include optics with grazing incidence mirrors, pinholes, and refractive lenses. Spatial, temporal, and spectral performances have been measured on x-ray tubes and UV lasers at CEA-DIF and at Physikalisch-Technische Bundesanstalt BESSY II synchrotron prior to be set on LMJ. GXI-1 and GXI-2 designs, metrology, and first experiments on LMJ are presented here.

OSTI ID:
22597086
Journal Information:
Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 3 Vol. 87; ISSN 0034-6748; ISSN RSINAK
Country of Publication:
United States
Language:
English

Similar Records

Laser Integration Line target diagnostics first results (invited)
Journal Article · Sun Oct 15 00:00:00 EDT 2006 · Review of Scientific Instruments · OSTI ID:20861391

Gas detectors for x-ray lasers
Journal Article · Thu May 01 00:00:00 EDT 2008 · Journal of Applied Physics · OSTI ID:21137252

Beamline for the characterization of radiation transfer standards ranging from 200 to 2000 eV
Journal Article · Sat Jul 01 00:00:00 EDT 1989 · Rev. Sci. Instrum.; (United States) · OSTI ID:5819175