Title: Spectroscopic determination of the composition of a 50 kV hydrogen diagnostic neutral beam

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.4961269· OSTI ID:22596411
; ;  [1];  [2]
  1. Department of Physics, University of Wisconsin–Madison, 1150 University Ave., Madison, Wisconsin 53706 (United States)
  2. Wheaton College, Wheaton, Illinois 60187 (United States)

A grating spectrometer with an electron multiplying charge-coupled device camera is used to diagnose a 50 kV, 5 A, 20 ms hydrogen diagnostic neutral beam. The ion source density is determined from Stark broadened H{sub β} emission and the spectrum of Doppler-shifted H{sub α} emission is used to quantify the fraction of ions at full, half, and one-third beam energy under a variety of operating conditions including fueling gas pressure and arc discharge current. Beam current is optimized at low-density conditions in the ion source while the energy fractions are found to be steady over most operating conditions.

OSTI ID:
22596411
Journal Information:
Review of Scientific Instruments, Vol. 87, Issue 11; Other Information: (c) 2016 Author(s); Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English

Similar Records

Grating spectrometer system for beam emission spectroscopy diagnostics using high-energy negative-ion-based neutral beam injection on LHD
Journal Article · Fri Oct 15 00:00:00 EDT 2010 · Review of Scientific Instruments · OSTI ID:22596411

Properties of an intense 50-kV neutral-beam injection system
Journal Article · Thu Apr 01 00:00:00 EST 1982 · Rev. Sci. Instrum.; (United States) · OSTI ID:22596411

Development of the plasma generator for a long pulse 10 x 40 neutral beam
Journal Article · Sat Nov 01 00:00:00 EST 1986 · Rev. Sci. Instrum.; (United States) · OSTI ID:22596411