Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Augmentation of Quick-EXAFS measurement facility at the energy scanning EXAFS beamline at INDUS-2 SRS

Journal Article · · AIP Conference Proceedings
DOI:https://doi.org/10.1063/1.4917862· OSTI ID:22490362
; ; ;  [1]
  1. Atomic and Molecular Physics Division, Bhabha Atomic Research Centre, Mumbai -400085 (India)
In this paper implementation of Quick-EXAFS data acquisition facility at the Energy Scanning EXAFS beamline(BL-09) at INDUS-2 synchrotron source, Indore is presented. By adopting a continuous-scan mode in the Double Crystal monochromator (DCM), a high signal-to-noise ratio is maintained and the acquisition time is reduced to few seconds. The quality of spectra and repeatability is checked by measuring standards. The present mode of data acquisition would enable EXAFS measurement for in-situ studies even in fluorescence mode.
OSTI ID:
22490362
Journal Information:
AIP Conference Proceedings, Journal Name: AIP Conference Proceedings Journal Issue: 1 Vol. 1665; ISSN APCPCS; ISSN 0094-243X
Country of Publication:
United States
Language:
English

Similar Records

Commissioning and first results of scanning type EXAFS beamline (BL-09) at INDUS-2 synchrotron source
Journal Article · Thu Apr 24 00:00:00 EDT 2014 · AIP Conference Proceedings · OSTI ID:22269533

Indus synchrotron source: A national facility
Journal Article · Tue Jun 05 00:00:00 EDT 2012 · AIP Conference Proceedings · OSTI ID:22004116

GIXRD measurements at EDXRD beamline at INDUS-2 synchrotron
Journal Article · Tue Jun 05 00:00:00 EDT 2012 · AIP Conference Proceedings · OSTI ID:22004098