Depth-resolved registration of transesophageal echo to x-ray fluoroscopy using an inverse geometry fluoroscopy system

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4935534· OSTI ID:22483058
 [1]; ; ;  [2];  [3];  [4]
  1. Department of Biomedical Engineering, University of Wisconsin-Madison, Madison, Wisconsin 53705 (United States)
  2. Department of Medical Physics, University of Wisconsin-Madison, Madison, Wisconsin 53705 (United States)
  3. Triple Ring Technologies, Inc., Newark, California 94560 (United States)
  4. Department of Medicine, University of Wisconsin-Madison, Madison, Wisconsin 53792 (United States)

Purpose: Image registration between standard x-ray fluoroscopy and transesophageal echocardiography (TEE) has recently been proposed. Scanning-beam digital x-ray (SBDX) is an inverse geometry fluoroscopy system designed for cardiac procedures. This study presents a method for 3D registration of SBDX and TEE images based on the tomosynthesis and 3D tracking capabilities of SBDX. Methods: The registration algorithm utilizes the stack of tomosynthetic planes produced by the SBDX system to estimate the physical 3D coordinates of salient key-points on the TEE probe. The key-points are used to arrive at an initial estimate of the probe pose, which is then refined using a 2D/3D registration method adapted for inverse geometry fluoroscopy. A phantom study was conducted to evaluate probe pose estimation accuracy relative to the ground truth, as defined by a set of coregistered fiducial markers. This experiment was conducted with varying probe poses and levels of signal difference-to-noise ratio (SDNR). Additional phantom and in vivo studies were performed to evaluate the correspondence of catheter tip positions in TEE and x-ray images following registration of the two modalities. Results: Target registration error (TRE) was used to characterize both pose estimation and registration accuracy. In the study of pose estimation accuracy, successful pose estimates (3D TRE < 5.0 mm) were obtained in 97% of cases when the SDNR was 5.9 or higher in seven out of eight poses. Under these conditions, 3D TRE was 2.32 ± 1.88 mm, and 2D (projection) TRE was 1.61 ± 1.36 mm. Probe localization error along the source-detector axis was 0.87 ± 1.31 mm. For the in vivo experiments, mean 3D TRE ranged from 2.6 to 4.6 mm and mean 2D TRE ranged from 1.1 to 1.6 mm. Anatomy extracted from the echo images appeared well aligned when projected onto the SBDX images. Conclusions: Full 6 DOF image registration between SBDX and TEE is feasible and accurate to within 5 mm. Future studies will focus on real-time implementation and application-specific analysis.

OSTI ID:
22483058
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 12 Vol. 42; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

TU-F-BRF-04: Registration of 3D Transesophageal Echocardiography and X-Ray Fluoroscopy Using An Inverse Geometry X-Ray System
Journal Article · Sun Jun 15 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22407816

Three-dimensional tracking of cardiac catheters using an inverse geometry x-ray fluoroscopy system
Journal Article · Tue Dec 14 23:00:00 EST 2010 · Medical Physics · OSTI ID:22096843

Left ventricular ejection fraction: Single-plane and multiplanar transesophageal echocardiography versus equilibrium gated-pool scintigraphy
Journal Article · Thu Jan 31 23:00:00 EST 1991 · Journal of Cardiothorac Vascular Anesthesia; (United States) · OSTI ID:5228522