Title: Three-dimensional tracking of cardiac catheters using an inverse geometry x-ray fluoroscopy system

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.3515463· OSTI ID:22096843
; ; ;  [1]
  1. Department of Medical Physics, University of Wisconsin-Madison, Madison, Wisconsin 53705 (United States)

Purpose: Scanning beam digital x-ray (SBDX) is an inverse geometry fluoroscopic system with high dose efficiency and the ability to perform continuous real-time tomosynthesis at multiple planes. This study describes a tomosynthesis-based method for 3D tracking of high-contrast objects and present the first experimental investigation of cardiac catheter tracking using a prototype SBDX system. Methods: The 3D tracking algorithm utilizes the stack of regularly spaced tomosynthetic planes that are generated by SBDX after each frame period (15 frames/s). Gradient-filtered versions of the image planes are generated, the filtered images are segmented into object regions, and then a 3D coordinate is calculated for each object region. Two phantom studies of tracking performance were conducted. In the first study, an ablation catheter in a chest phantom was imaged as it was pulled along a 3D trajectory defined by a catheter sheath (10, 25, and 50 mm/s pullback speeds). SBDX tip tracking coordinates were compared to the 3D trajectory of the sheath as determined from a CT scan of the phantom after the registration of the SBDX and CT coordinate systems. In the second study, frame-to-frame tracking precision was measured for six different catheter configurations as a function of image noise level (662-7625 photons/mm{sup 2} mean detected x-ray fluence at isocenter). Results: During catheter pullbacks, the 3D distance between the tracked catheter tip and the sheath centerline was 1.0{+-}0.8 mm (mean {+-}one standard deviation). The electrode to centerline distances were comparable to the diameter of the catheter tip (2.3 mm), the confining sheath (4 mm outside diameter), and the estimated SBDX-to-CT registration error ({+-}0.7 mm). The tip position was localized for all 332 image frames analyzed and 83% of tracked positions were inside the 3D sheath volume derived from CT. The pullback speeds derived from the catheter trajectories were within 5% of the programed pullback speeds. The tracking precision of ablation and diagnostic catheter tips ranged from {+-}0.2 mm at the highest image fluence to {+-}0.9 mm at the lowest fluence. Tracking precision depended on image fluence, the size of the tracked catheter electrode, and the contrast of the electrode. Conclusions: High speed multiplanar tomosynthesis with an inverse geometry x-ray fluoroscopy system enables 3D tracking of multiple high-contrast objects at the rate of fluoroscopic imaging. The SBDX system is capable of tracking electrodes in standard cardiac catheters with approximately 1 mm accuracy and precision.

OSTI ID:
22096843
Journal Information:
Medical Physics, Vol. 37, Issue 12; Other Information: (c) 2010 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English