Estimating pediatric entrance skin dose from digital radiography examination using DICOM metadata: A quality assurance tool

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4918324· OSTI ID:22413558

Purpose: To develop an automated methodology to estimate patient examination dose in digital radiography (DR) imaging using DICOM metadata as a quality assurance (QA) tool. Methods: Patient examination and demographical information were gathered from metadata analysis of DICOM header data. The x-ray system radiation output (i.e., air KERMA) was characterized for all filter combinations used for patient examinations. Average patient thicknesses were measured for head, chest, abdomen, knees, and hands using volumetric images from CT. Backscatter factors (BSFs) were calculated from examination kVp. Patient entrance skin air KERMA (ESAK) was calculated by (1) looking up examination technique factors taken from DICOM header metadata (i.e., kVp and mA s) to derive an air KERMA (k{sub air}) value based on an x-ray characteristic radiation output curve; (2) scaling k{sub air} with a BSF value; and (3) correcting k{sub air} for patient thickness. Finally, patient entrance skin dose (ESD) was calculated by multiplying a mass–energy attenuation coefficient ratio by ESAK. Patient ESD calculations were computed for common DR examinations at our institution: dual view chest, anteroposterior (AP) abdomen, lateral (LAT) skull, dual view knee, and bone age (left hand only) examinations. Results: ESD was calculated for a total of 3794 patients; mean age was 11 ± 8 yr (range: 2 months to 55 yr). The mean ESD range was 0.19–0.42 mGy for dual view chest, 0.28–1.2 mGy for AP abdomen, 0.18–0.65 mGy for LAT view skull, 0.15–0.63 mGy for dual view knee, and 0.10–0.12 mGy for bone age (left hand) examinations. Conclusions: A methodology combining DICOM header metadata and basic x-ray tube characterization curves was demonstrated. In a regulatory era where patient dose reporting has become increasingly in demand, this methodology will allow a knowledgeable user the means to establish an automatable dose reporting program for DR and perform patient dose related QA testing for digital x-ray imaging.

OSTI ID:
22413558
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 5 Vol. 42; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

Reference dosimetry during diagnostic CT examination using XR-QA radiochromic film model
Journal Article · Thu Sep 15 00:00:00 EDT 2011 · Medical Physics · OSTI ID:22098616

A Monte Carlo estimation of effective dose in chest tomosynthesis
Journal Article · Mon Dec 14 23:00:00 EST 2009 · Medical Physics · OSTI ID:22098480

Dosimetric Quantities for Computed Tomography Examinations of Paediatric Patients on the Thoracic and Abdominal Regions
Journal Article · Mon Dec 06 23:00:00 EST 2010 · AIP Conference Proceedings · OSTI ID:21506979