Accuracy of an infrared marker-based patient positioning system (ExacTrac®) for stereotactic body radiotherapy in localizing the planned isocenter using fiducial markers

Journal Article · · AIP Conference Proceedings
DOI:https://doi.org/10.1063/1.4901367· OSTI ID:22390527
 [1]; ; ; ; ; ;  [2]
  1. Medical Physics Graduate Programme, Universidad Autónoma del Estado de México, Facultad de Medicina, Paseo Tollocan esquina Jesús Carranza Colonia Moderna de la Cruz, C.P. 50180, Toluca, Estado de México (Mexico)
  2. Centro de Cáncer ABC, The American British Cowdray Medical Center I.A.P. (CMABC), Calle Sur 136, no. 116, Colonia las Américas, C.P. 01120, México, D.F. (Mexico)

Stereotactic Body Radiation Therapy (SBRT) requires a controlled immobilization and position monitoring of patient and target. The purpose of this work is to analyze the performance of the imaging system ExacTrac® (ETX) using infrared and fiducial markers. Materials and methods: In order to assure the accuracy of isocenter localization, a Quality Assurance procedure was applied using an infrared marker-based positioning system. Scans were acquired of an inhouse-agar gel and solid water phantom with infrared spheres. In the inner part of the phantom, three reference markers were delineated as reference and one pellet was place internally; which was assigned as the isocenter. The iPlan® RT Dose treatment planning system. Images were exported to the ETX console. Images were acquired with the ETX to check the correctness of the isocenter placement. Adjustments were made in 6D the reference markers were used to fuse the images. Couch shifts were registered. The procedure was repeated for verification purposes. Results: The data recorded of the verifications in translational and rotational movements showed averaged 3D spatial uncertainties of 0.31 ± 0.42 mm respectively 0.82° ± 0.46° in the phantom and the first correction of these uncertainties were of 1.51 ± 1.14 mm respectively and 1.37° ± 0.61°. Conclusions: This study shows a high accuracy and repeatability in positioning the selected isocenter. The ETX-system for verifying the treatment isocenter position has the ability to monitor the tracing position of interest, making it possible to be used for SBRT positioning within uncertainty ≤1mm.

OSTI ID:
22390527
Journal Information:
AIP Conference Proceedings, Journal Name: AIP Conference Proceedings Journal Issue: 1 Vol. 1626; ISSN 0094-243X; ISSN APCPCS
Country of Publication:
United States
Language:
English

Similar Records

EXACTRAC x-ray and beam isocenters--What's the difference?
Journal Article · Thu Mar 15 00:00:00 EDT 2012 · Medical Physics · OSTI ID:22098786

Assessment of Spatial Uncertainties in the Radiotherapy Process With the Novalis System
Journal Article · Thu Oct 01 00:00:00 EDT 2009 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:21282060

SU-E-T-19: A New End-To-End Test Method for ExacTrac for Radiation and Plan Isocenter Congruence
Journal Article · Sun Jun 01 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22339798