Assessment of Spatial Uncertainties in the Radiotherapy Process With the Novalis System

Journal Article · · International Journal of Radiation Oncology, Biology and Physics
 [1];  [2];  [3]; ; ;  [1]
  1. Nagoya Radiosurgery Center, Nagoya Kyoritsu Hospital, Nagoya (Japan)
  2. Graduate school of Medical Science, Nagoya University, Nagoya (Japan)
  3. School of Health Sciences, Gifu University of Medical Science, Gifu (Japan)

Purpose: The purpose of this study was to evaluate the accuracy of a new version of the ExacTrac X-ray (ETX) system with statistical analysis retrospectively in order to determine the tolerance of systematic components of spatial uncertainties with the Novalis system. Methods and Materials: Three factors of geometrical accuracy related to the ETX system were evaluated by phantom studies. First, location dependency of the detection ability of the infrared system was evaluated. Second, accuracy of the automated calculation by the image fusion algorithm in the patient registration software was evaluated. Third, deviation of the coordinate scale between the ETX isocenter and the mechanical isocenter was evaluated. From the values of these examinations and clinical experiences, the total spatial uncertainty with the Novalis system was evaluated. Results: As to the location dependency of the detection ability of the infrared system, the detection errors between the actual position and the detected position were 1% in translation shift and 0.1{sup o} in rotational angle, respectively. As to the accuracy of patient verification software, the repeatability and the coincidence of the calculation value by image fusion were good when the contrast of the X-ray image was high. The deviation of coordinates between the ETX isocenter and the mechanical isocenter was 0.313 {+-} 0.024 mm, in a suitable procedure. Conclusions: The spatial uncertainty will be less than 2 mm when suitable treatment planning, optimal patient setup, and daily quality assurance for the Novalis system are achieved in the routine workload.

OSTI ID:
21282060
Journal Information:
International Journal of Radiation Oncology, Biology and Physics, Journal Name: International Journal of Radiation Oncology, Biology and Physics Journal Issue: 2 Vol. 75; ISSN IOBPD3; ISSN 0360-3016
Country of Publication:
United States
Language:
English

Similar Records

Accuracy of an infrared marker-based patient positioning system (ExacTrac®) for stereotactic body radiotherapy in localizing the planned isocenter using fiducial markers
Journal Article · Thu Nov 06 23:00:00 EST 2014 · AIP Conference Proceedings · OSTI ID:22390527

SU-E-J-48: Imaging Origin-Radiation Isocenter Coincidence for Linac-Based SRS with Novalis Tx
Journal Article · Mon Jun 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:22494071

EXACTRAC x-ray and beam isocenters--What's the difference?
Journal Article · Thu Mar 15 00:00:00 EDT 2012 · Medical Physics · OSTI ID:22098786