Title: Investigation of potential polyatomic interferences on uranium isotope ratio measurements for the LS-APGD-Orbitrap MS system

Journal Article · · Microchemical Journal

The determination of actinide (e.g., U and Pu) content and isotopics is of importance to the nuclear forensics and safeguards communities. However, in the analysis of environmental samples, such as those collected by the International Atomic Energy Agency, uranium measurements can be complicated by isobaric interferences from polyatomic variants of heavy elements (e.g., Pb). This leads to complex, time-consuming sample manipulations (i.e., separations) before determining isotope ratios. An alternative strategy to sample pretreatment is to use high-resolution mass spectrometric platforms during the analysis to fully resolve the uranium isotopes from potential polyatomic interferences, negating the need for prior chemical separation. The liquid sampling - atmospheric pressure glow discharge (LS-APGD) coupled with an Orbitrap mass spectrometer provides a high-resolution (>70,000 m/Δm at m/z 200) inorganic mass spectrometry platform. Further, as a demonstration of the power of this instrumental platform, and indeed, the high-resolution approach in general, uranium isotope ratios were determined in the presence of elemental impurities (e.g., Pb, Pt, Ta, W) commonly encountered with environmental sample swipe analysis, without any prior treatment. Even at elemental impurity concentrations of 1000–5000× relative to uranium, no interference was observed with the 235U or 238U signal. In addition, the 235U/238U isotope ratio for the samples with the concomitants present are within 2 standard deviations of the values obtained without their addition, indicating that these impurities do not impact the determined uranium isotope ratio. These findings represent a significant first step in leveraging the high resolution of the LS-APGD-Orbitrap-MS to overcome isobaric and molecular interferences instead of relying on chemical separations.

Research Organization:
Oak Ridge National Laboratory (ORNL), Oak Ridge, TN (United States)
Sponsoring Organization:
USDOE National Nuclear Security Administration (NNSA)
Grant/Contract Number:
AC05-00OR22725
OSTI ID:
2224194
Journal Information:
Microchemical Journal, Vol. 196, Issue 1; ISSN 0026-265X
Publisher:
ElsevierCopyright Statement
Country of Publication:
United States
Language:
English

References (43)

Assessment of plutonium measurement in the femtogram range by ICP-MS; correction from interfering polyatomic species journal January 2008
Automated Separation of Uranium and Plutonium from Environmental Swipe Samples for Multiple Collector Inductively Coupled Plasma Mass Spectrometry journal June 2018
Coupling of the Liquid Sampling–Atmospheric Pressure Glow Discharge to Orbitrap Mass Analyzers for Uranium Isotope Ratio Analysis: Evolution of the Methodology and Implications to the Field journal January 2023
Rapid and automated separation of uranium ore concentrates for trace element analysis by inductively coupled plasma – optical emission spectroscopy/triple quadrupole mass spectrometry journal May 2021
Roles of collisional dissociation modalities on spectral composition and isotope ratio measurement performance of the liquid sampling – atmospheric pressure glow discharge / orbitrap mass spectrometer coupling journal June 2021
Using CO2 Reactions to Achieve Mass-spectrometric Discrimination in Simultaneous Plutonium-isotope Speciation with Inductively Coupled Plasma–Tandem Mass Spectrometry journal July 2022
Improved Uranium Isotope Ratio Analysis in Liquid Sampling–Atmospheric Pressure Glow Discharge/Orbitrap FTMS Coupling through the Use of an External Data Acquisition System journal April 2021
Liquid sampling-atmospheric pressure glow discharge optical emission spectroscopy detection of laser ablation produced particles: A feasibility study journal October 2012
Initial Characterization and Optimization of the Liquid Sampling-Atmospheric Pressure Glow Discharge Ionization Source Coupled to an Orbitrap Mass Spectrometer for the Determination of Plutonium journal July 2023
Exploration of ICP platforms for measuring elemental impurities in uranium ore concentrates journal September 2020
Liquid sampling-atmospheric pressure glow discharge (LS-APGD) microplasmas for diverse spectrochemical analysis applications journal January 2017
Mass shifts and local space charge effects observed in the quadrupole ion trap at higher resolution journal May 1995
A novel method for measuring ultra-trace levels of U and Th in Au, Pt, Ir, and W matrices using ICP-QQQ-MS employing an O2 reaction gas journal January 2020
Initial Benchmarking of the Liquid Sampling-Atmospheric Pressure Glow Discharge-Orbitrap System Against Traditional Atomic Mass Spectrometry Techniques for Nuclear Applications journal October 2018
Strengthening IAEA safeguards through environmental sampling and analysis journal June 1998
Ambient desorption/ionization mass spectrometry using a liquid sampling–atmospheric glow discharge (LS-APGD) ionization source journal July 2013
Total evaporation method for uranium isotope-amount ratio measurements journal January 2013
Evolution of Collision/Reaction Cell Technology — Advent of Triple Quadrupole ICP-MS journal January 2018
Direct Uranium Isotopic Analysis of Swipe Surfaces by Microextraction-ICP-MS journal July 2021
The use of carbon dioxide as the reaction cell gas for the separation of uranium and plutonium in quadrupole inductively coupled plasma mass spectrometry (ICP-MS) for nuclear forensic samples journal August 2018
Liquid Sampling-Atmospheric Pressure Glow Discharge Ionization Source for Elemental Mass Spectrometry journal April 2011
Coupling of Laser Ablation and the Liquid Sampling-Atmospheric Pressure Glow Discharge Plasma for Simultaneous, Comprehensive Mapping: Elemental, Molecular, and Spatial Analysis journal August 2020
Observation of ion coalescence in Orbitrap Fourier transform mass spectrometry: Ion coalescence in Orbitrap FTMS journal June 2012
A new method for determining 236U/238U isotope ratios in environmental samples by means OF ICP-MS/MS journal January 2020
Preparation and certification of the uranium nitrate solution reference materials series IRMM-2019 to IRMM-2029 for the isotopic composition journal September 2018
Pushing Limits of ICP–MS/MS for the Determination of Ultralow 236U/238U Isotope Ratios journal May 2020
Mass spectrometric analysis for nuclear safeguards journal January 2015
Development, validation and verification of an ICP-MS procedure for a multi-element analysis of uranium ore concentrates journal January 2017
Conceptual Demonstration of Ambient Desorption-Optical Emission Spectroscopy Using a Liquid Sampling-Atmospheric Pressure Glow Discharge Microplasma Source journal May 2016
Analysis of solid uranium particulates on cotton swipes with an automated microextraction-ICP-MS system journal January 2022
Direct isotopic analysis of solid uranium particulates on cotton swipes by microextraction-ICP-MS journal May 2022
Optimization of uranium and plutonium separations using TEVA and UTEVA cartridges for MC-ICP-MS analysis of environmental swipe samples journal June 2019
Sector field mass spectrometers in ICP-MS journal November 1998
Improved uranium isotopic ratio determinations for the liquid sampling-atmospheric pressure glow discharge orbitrap mass spectrometer by use of moving average processing journal January 2022
Peer Reviewed: Strengthened Nuclear Safeguards journal January 2002
Concomitant ion effects on isotope ratio measurements with liquid sampling – atmospheric pressure glow discharge ion source Orbitrap mass spectrometry journal January 2018
Radionuclide analysis using collision–reaction cell ICP-MS technology: a review journal January 2020
Resolving Severe Elemental Isobaric Interferences with a Combined Atomic and Molecular Ionization Source–Orbitrap Mass Spectrometry Approach: The 87Sr and 87Rb Geochronology Pair journal August 2021
Direct analysis of cotton swipes for plutonium isotope determination by microextraction-ICP-MS journal January 2021
Rapid Determination of Uranium Isotopic Abundance from Cotton Swipes: Direct Extraction via a Planar Surface Reader and Coupling to a Microplasma Ionization Source journal June 2020
A multi-electrode glow discharge ionization source for atomic and molecular mass spectrometry journal January 2020
Strategies to develop methods using ion-molecule reactions in a quadrupole reaction cell to overcome spectral overlaps in inductively coupled plasma mass spectrometry journal January 2006
Polyatomic interferences on high precision uranium isotope ratio measurements by MC-ICP-MS: applications to environmental sampling for nuclear safeguards journal September 2015