Title: Safety and Efficacy of the Gunther Tulip Retrievable Vena Cava Filter: Midterm Outcomes

Journal Article · · Cardiovascular and Interventional Radiology
; ; ; ;  [1]
  1. Dartmouth-Hitchcock Medical Center, Department of Radiology, Section of Vascular and Interventional Radiology (United States)

PurposeTo evaluate of the medium-term integrity, efficacy, and complication rate associated with the Gunther Tulip vena cava filter.MethodsA retrospective study was performed of 369 consecutive patients who had infrarenal Gunther Tulip inferior vena cava filters placed over a 5-year period. The mean patient age was 61.8 years, and 59 % were men. Venous thromboembolic disease and a contraindication to or complication of anticoagulation were the indications for filter placement in 86 % of patients; 14 % were placed for prophylaxis in patients with a mean of 2.3 risk factors. Follow-up was obtained by review of medical and radiologic records.ResultsMean clinical follow-up was 780 days. New or recurrent pulmonary embolus occurred in 12 patients (3.3 %). New or recurrent deep-vein thrombosis occurred in 53 patients (14.4 %). There were no symptomatic fractures, migrations, or caval perforations. Imaging follow-up in 287 patients (77.8 %) at a mean of 731 days revealed a single (0.3 %) asymptomatic fracture, migration greater than 2 cm in 36 patients (12.5 %), and no case of embolization. Of 122 patients with CT scans, asymptomatic perforations were identified in 53 patients (43.4 %) at a mean 757 days.ConclusionThe Gunther Tulip filter was safe and effective at 2-year follow-up. Complication rates were similar to those reported for permanent inferior vena cava filters.

OSTI ID:
22121738
Journal Information:
Cardiovascular and Interventional Radiology, Vol. 36, Issue 4; Other Information: Copyright (c) 2013 Springer Science+Business Media New York and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE); Country of input: International Atomic Energy Agency (IAEA); ISSN 0174-1551
Country of Publication:
United States
Language:
English

Similar Records

Evaluation of Retrievability of the Gunther Tulip Vena Cava Filter
Journal Article · Sun Apr 15 00:00:00 EDT 2007 · Cardiovascular and Interventional Radiology · OSTI ID:22121738

Gunther Tulip Retrievable Inferior Vena Caval Filters: Indications, Efficacy, Retrieval, and Complications
Journal Article · Thu Feb 15 00:00:00 EST 2007 · Cardiovascular and Interventional Radiology · OSTI ID:22121738

Gunther Tulip Inferior Vena Cava Filter Placement During Treatment for Deep Venous Thrombosis of the Lower Extremity
Journal Article · Sun May 15 00:00:00 EDT 2005 · Cardiovascular and Interventional Radiology · OSTI ID:22121738