Progress in obtaining an absolute calibration of a total deuterium-tritium neutron yield diagnostic based on copper activation
- Sandia National Laboratories, Albuquerque, New Mexico 87185 (United States)
- Chemical and Nuclear Engineering Department, University of New Mexico, Albuquerque, New Mexico 87131 (United States)
The 350-keV Cockroft-Walton accelerator at Sandia National laboratory's Ion Beam facility is being used to calibrate absolutely a total DT neutron yield diagnostic based on the {sup 63}Cu(n,2n){sup 62}Cu({beta}+) reaction. These investigations have led to first-order uncertainties approaching 5% or better. The experiments employ the associated-particle technique. Deuterons at 175 keV impinge a 2.6 {mu}m thick erbium tritide target producing 14.1 MeV neutrons from the T(d,n){sup 4}He reaction. The alpha particles emitted are measured at two angles relative to the beam direction and used to infer the neutron flux on a copper sample. The induced {sup 62}Cu activity is then measured and related to the neutron flux. This method is known as the F-factor technique. Description of the associated-particle method, copper sample geometries employed, and the present estimates of the uncertainties to the F-factor obtained are given.
- OSTI ID:
- 22093880
- Journal Information:
- Review of Scientific Instruments, Vol. 83, Issue 10; Other Information: (c) 2012 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
- Country of Publication:
- United States
- Language:
- English
Similar Records
Absolute calibration of a total yield indium activation detector for DD and DT neutrons
Absolute calibration of a total yield indium activation detector for DD and DT neutrons