Measuring the absolute deuterium-tritium neutron yield using the magnetic recoil spectrometer at OMEGA and the NIF

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.4738657· OSTI ID:22093879
; ; ; ; ;  [1]; ; ; ; ;  [2]; ; ; ; ; ; ; ; ;  [3] more »; « less
  1. Plasma Science and Fusion Center, MIT, Cambridge, Massachusetts 02139 (United States)
  2. Laboratory for Laser Energetics, UR, Rochester, New York 14623 (United States)
  3. Lawrence Livermore National Laboratory, Livermore, California 94550 (United States)
A magnetic recoil spectrometer (MRS) has been installed and extensively used on OMEGA and the National Ignition Facility (NIF) for measurements of the absolute neutron spectrum from inertial confinement fusion implosions. From the neutron spectrum measured with the MRS, many critical implosion parameters are determined including the primary DT neutron yield, the ion temperature, and the down-scattered neutron yield. As the MRS detection efficiency is determined from first principles, the absolute DT neutron yield is obtained without cross-calibration to other techniques. The MRS primary DT neutron measurements at OMEGA and the NIF are shown to be in excellent agreement with previously established yield diagnostics on OMEGA, and with the newly commissioned nuclear activation diagnostics on the NIF.
OSTI ID:
22093879
Journal Information:
Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 10 Vol. 83; ISSN 0034-6748; ISSN RSINAK
Country of Publication:
United States
Language:
English