The neutron spectrum produced by deuterium-tritium (DT) inertial confinement fusion implosions contains a wealth of information about implosion performance including the DT yield, ion-temperature, and areal-density. The Magnetic Recoil Spectrometer (MRS) has been used at both the OMEGA laser facility and the National Ignition Facility (NIF) to measure the absolute neutron spectrum from 3 to 30 MeV at OMEGA and 3 to 36 MeV at the NIF. These measurements have been used to diagnose the performance of cryogenic target implosions to unprecedented accuracy. Interpretation of MRS data requires a detailed understanding of the MRS response and background. This paper describes ab initio characterization of the system involving Monte Carlo simulations of the MRS response in addition to the commission experiments for in situ calibration of the systems on OMEGA and the NIF.
- OSTI ID:
- 22105506
- Journal Information:
- Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 4 Vol. 84; ISSN 0034-6748; ISSN RSINAK
- Country of Publication:
- United States
- Language:
- English
Similar Records
The magnetic recoil spectrometer for measurements of the absolute neutron spectrum at OMEGA and the NIF
The coincidence counting technique for orders of magnitude background reduction in data obtained with the magnetic recoil spectrometer at OMEGA and the NIF
Measuring the absolute DT neutron yield using the Magnetic Recoil Spectrometer at OMEGA and the NIF
Journal Article
·
Wed Apr 17 20:00:00 EDT 2013
· Review of Scientific Instruments
·
OSTI ID:1172369
The coincidence counting technique for orders of magnitude background reduction in data obtained with the magnetic recoil spectrometer at OMEGA and the NIF
Journal Article
·
Fri Jul 15 00:00:00 EDT 2011
· Review of Scientific Instruments
·
OSTI ID:22075521
Measuring the absolute DT neutron yield using the Magnetic Recoil Spectrometer at OMEGA and the NIF
Conference
·
Thu May 03 00:00:00 EDT 2012
·
OSTI ID:1043652
Related Subjects
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
70 PLASMA PHYSICS AND FUSION TECHNOLOGY
COMPUTERIZED SIMULATION
DEUTERIUM
ELECTRON TEMPERATURE
INERTIAL CONFINEMENT
ION TEMPERATURE
MONTE CARLO METHOD
NEUTRON SPECTRA
NEUTRON SPECTROMETERS
PLASMA
PLASMA DIAGNOSTICS
THERMONUCLEAR REACTORS
TRITIUM
US NATIONAL IGNITION FACILITY
70 PLASMA PHYSICS AND FUSION TECHNOLOGY
COMPUTERIZED SIMULATION
DEUTERIUM
ELECTRON TEMPERATURE
INERTIAL CONFINEMENT
ION TEMPERATURE
MONTE CARLO METHOD
NEUTRON SPECTRA
NEUTRON SPECTROMETERS
PLASMA
PLASMA DIAGNOSTICS
THERMONUCLEAR REACTORS
TRITIUM
US NATIONAL IGNITION FACILITY