skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Routine {sup 18}F{sup -} production with 180 {mu}A to 200 {mu}A target beam current on a GE PETtrace 800 cyclotron

Journal Article · · AIP Conference Proceedings
DOI:https://doi.org/10.1063/1.4773942· OSTI ID:22075954
; ; ; ; ; ; ;  [1]
  1. Department of PET and Nuclear Medicine, Royal Prince Alfred Hospital, Sydney, NSW (Australia)

Through upgrades and enhancements, which are now standard on current PETtrace 800 cyclotrons, a GE PETtrace cyclotron installed in 2002 has demonstrated reliable routine [{sup 18}F]FDG production at total target beam currents of 180 {mu}A without detriment to saturation and [{sup 18}F]FDG yields. Routine production at 200 {mu}A has been achieved and its evaluation is continuing. Self-shielded target using W/Cu alloy for the target body afforded a reduction in dose rate from the Havar foils by a factor of {approx} 8-10, reducing dose from the targets and need for removing targets during maintenance. The main activation product in the shield is {sup 187}W (T1/2 24 h). The {sup 60}Co ((T1/2 5.3 y) activation is about 250 times less at 24 h post EOB and is not considered a major issue despite its long half-life.

OSTI ID:
22075954
Journal Information:
AIP Conference Proceedings, Vol. 1509, Issue 1; Conference: 14. international workshop on targetry and target chemistry, Playa del Carmen (Mexico), 26-29 Aug 2012; Other Information: (c) 2012 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-243X
Country of Publication:
United States
Language:
English