Stabilization of sample temperature in a surface-science vacuum chamber to 0.03 K and quartz-crystal microbalance frequency to 0.06 Hz over 0.5 h

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.3488368· OSTI ID:22058633
 [1]
  1. Department of Physics and Astronomy, Trent University, Peterborough, Ontario K9J7B8 (Canada)
Improvements have been made to a high-stability quartz-crystal microbalance for use in a typical surface-science, ultrahigh vacuum chamber, with a frequency stability of one part in 10{sup 8} (0.06 Hz) over 0.5 h. This gives a resolution equivalent to 2% of an atomic monolayer of oxygen over 0.5 h. The quartz-crystal microbalance (QCM) crystal can be rotated to different surface-analysis positions in the chamber. These characteristics open up the combination of surface and bulk adsorption studies on the same sample without transferring the sample to another chamber. To accomplish this, it was necessary to stabilize the sample temperature to {+-}0.03 K over several hours. The oscillator performance is illustrated by the uptake of oxygen by a gold-plated QCM crystal.
OSTI ID:
22058633
Journal Information:
Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 10 Vol. 81; ISSN 0034-6748; ISSN RSINAK
Country of Publication:
United States
Language:
English

Similar Records

Angular distribution of sputtered particles measured by quartz crystal microbalances and Auger electron spectroscopy
Journal Article · Sun Mar 02 23:00:00 EST 1986 · Appl. Phys. Lett.; (United States) · OSTI ID:6102031

Troubleshoot Quartz Crystal Microbalance Measurements Data
Technical Report · Sun Dec 04 19:00:00 EST 2011 · OSTI ID:1033737

Quartz microbalance device for transfer into ultrahigh vacuum systems
Journal Article · Wed Oct 15 00:00:00 EDT 2008 · Review of Scientific Instruments · OSTI ID:21266460