Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Development and performance analysis of a standard orifice flow calibration system

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.3176469· OSTI ID:22053530
; ;  [1]
  1. NINVAST, National Centre for Physics Complex, Shahdara Valley Road, P.O. Box 3125, Islamabad (Pakistan)

An orifice flow system has been developed indigenously which is a primary standard for the calibration of high vacuum gauges in the range of 10{sup -3}-10{sup -6} mbar. It consists of a constant-volume flow meter, designed to generate a known flow rate of a particular gas to the vacuum chamber. This chamber is partitioned into two parts by a plate with an orifice of calculable conductance. The pressures, generated by this standard system, are compared with those of secondary standard, namely, spinning rotor gauge. Different uncertainties and correction factors are calculated. The maximum percentage deviation is from 1.16% to 0.97%. The combined uncertainties over the entire calibration range of the standard system are in the range of 4.0x10{sup -6} mbar.

OSTI ID:
22053530
Journal Information:
Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 7 Vol. 80; ISSN 0034-6748; ISSN RSINAK
Country of Publication:
United States
Language:
English

Similar Records

An automated vacuum gauge calibration system
Conference · Tue Mar 31 23:00:00 EST 1998 · OSTI ID:654189

Low-range flowmeters for use with vacuum and leak standards
Journal Article · Fri May 01 00:00:00 EDT 1987 · J. Vac. Sci. Technol., A; (United States) · OSTI ID:6876243

USE OF A STANDARD ORIFICE IN THE CALIBRATION OF VACUUM GAUGES
Technical Report · Mon May 01 00:00:00 EDT 1961 · OSTI ID:4065614