Title: Analysis of order contamination from toroidal grating monochromator of Indus-1 reflectivity beamline

Journal Article · · AIP Conference Proceedings
DOI:https://doi.org/10.1063/1.4710104· OSTI ID:22004106
; ;  [1]
  1. X-ray Optics Section, Indus Synchrotron Utilization Division Raja Ramanna Centre for Advanced Technology, Indore 452 013 (India)

The synchrotron radiation spectra of toroidal grating monochromators (TGM) often contaminated by significant amounts of higher orders. Quantitative estimation of order contamination is essential in order to perform quantitative spectroscopy measurements. At Indus-1 reflectivity beamline higher order contaminations from TGM monochromator are experimentally determined using a reflection grating analyzer. It is found that the order contamination in wavelength range of 50-200 A is less than 15%. Use of edge filters reduces the order contamination level below 5% in this wavelength range.

OSTI ID:
22004106
Journal Information:
AIP Conference Proceedings, Vol. 1447, Issue 1; Conference: 56. DAE solid state physics symposium 2011, Kattankulathur, Tamilnadu (India), 19-23 Dec 2011; Other Information: (c) 2012 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-243X