Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Evaluation of Patient Doses from Verification Techniques in Image-Guided Radiotherapy (IGRT)

Journal Article · · AIP Conference Proceedings
DOI:https://doi.org/10.1063/1.3295647· OSTI ID:21366911
 [1]; ;  [2];  [3]; ;  [4]
  1. Faculty of Nuclear Sciences and Physical Engineering, Czech Technical University in Prague, Brehova 7, 115 19 Prague 1 (Czech Republic)
  2. National Radiation Protection Institute, Bartoskova 28, 140 00, Prague 4 (Czech Republic)
  3. Faculty Hospital Motol, Vuvalu 84, 150 06, Prague 5 (Czech Republic)
  4. Regional Hospital Liberec, Husova 10, 460 63, Liberec (Czech Republic)

The purpose of this work was an evaluation of organ doses and effective doses from kilovoltage (kV) cone-beam CT (CBCT) scans and from pairs of orthogonal kV image projections for two different treatment sites (a head and a pelvis). Measurements of organ doses were performed in an anthropomorphic rando phantom by means of thermoluminescent dosimeters (TLDs). Irradiations were performed using on-board imager (OBI) and X-ray volume imaging (XVI) systems.

OSTI ID:
21366911
Journal Information:
AIP Conference Proceedings, Journal Name: AIP Conference Proceedings Journal Issue: 1 Vol. 1204; ISSN APCPCS; ISSN 0094-243X
Country of Publication:
United States
Language:
English