Flow Velocities After Carotid Artery Stenting: Impact of Stent Design. A Fluid Dynamics Study in a Carotid Artery Model with Laser Doppler Anemometry

Journal Article · · Cardiovascular and Interventional Radiology
; ;  [1]; ;  [2]; ;  [3]; ;  [4];  [1]
  1. Technical University of Munich, Department of Interventional Radiology, Klinikum rechts der Isar (Germany)
  2. Technical University of Munich, Department of Vascular Surgery, Klinikum rechts der Isar (Germany)
  3. Abbott Company, Vascular Devices (Switzerland)
  4. University of Applied Science Munich, Laboratory for Fluid Mechanics (Germany)
Purpose. To study the influence of a newly developed membrane stent design on flow patterns in a physiologic carotid artery model. Methods. Three different stents were positioned in silicone models of the carotid artery: a stainless steel stent (Wall-stent), a nitinol stent (SelfX), and a nitinol stent with a semipermeable membrane (MembraX). To increase the contact area of the membrane with the vessel wall, another MembranX model was modified at the outflow tract. The membrane consists of a biocompatible silicone-polyurethane copolymer (Elast-Eon) with a pore size of 100 {mu}m. All stents were deployed across the bifurcation and the external carotid artery origin. Flow velocity measurements were performed with laser Doppler anemometry (LDA), using pulsatile flow conditions (Re = 220; flow 0.39 l/min; flow rate ratio ICA:ECA = 70:30) in hemodynamically relevant cross-sections. The hemodynamic changes were analyzed by comparing velocity fluctuations of corresponding flow profiles. Results. The flow rate ratio ICA:ECA shifted significantly from 70/30 to 73.9/26.1 in the MembraX and remained nearly unchanged in the SelfX and Wallstent. There were no changes in the flow patterns at the inflow proximal to the stents. In the stent no relevant changes were found in the SelfX. In the Wallstent the separation zone shifted from the orifice of the ICA to the distal end of the stent. Four millimeters distal to the SelfX and the Wallstent the flow profile returned to normal. In the MembraX an increase in the central slipstreams was found with creation of a flow separation distal to the stent. With a modification of the membrane this flow separation vanished. In the ECA flow disturbances were seen at the inner wall distal to the stent struts in the SelfX and the Wallstent. With the MembraX a calming of flow could be observed in the ECA with a slight loss of flow volume. Conclusions. Stent placement across the carotid artery bifurcation induces alterations of the physiologic flow behavior. Depending on the stent design the flow alterations are located in different regions. All the stents tested were suitable for the carotid bifurcation. The MembraX prototype has shown promising hemodynamic properties ex vivo.
OSTI ID:
21088166
Journal Information:
Cardiovascular and Interventional Radiology, Journal Name: Cardiovascular and Interventional Radiology Journal Issue: 1 Vol. 28; ISSN 0174-1551; ISSN CAIRDG
Country of Publication:
United States
Language:
English

Similar Records

Extracranial Carotid Artery Stenting in Surgically High-Risk Patients Using the Carotid Wallstent Endoprosthesis:Midterm Clinical and Ultrasound Follow-Up Results
Journal Article · Fri Aug 15 00:00:00 EDT 2003 · Cardiovascular and Interventional Radiology · OSTI ID:21088369

Carotid Artery Stenting Using a Novel Self-Expanding Braided Nickel-Titanium Stent: Feasibility and Safety Porcine Trial
Journal Article · Tue Sep 15 00:00:00 EDT 2009 · Cardiovascular and Interventional Radiology · OSTI ID:21429114

Neointimal Hyperplasia in Low-Profile Nitinol Stents, Palmaz Stents, and Wallstents: A Comparative Experimental Study
Journal Article · Mon Apr 15 00:00:00 EDT 1996 · Cardiovascular and Interventional Radiology · OSTI ID:21080520