Title: The effective point of measurement of ionization chambers and the build-up anomaly in MV x-ray beams

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.2839329· OSTI ID:21036187
; ;  [1]
  1. Ionizing Radiation Standards, Institute for National Measurement Standards, National Research Council of Canada, Ottawa, Ontario K1A 0R6 (Canada)

A precision experimental investigation of the effective point of measurement (EPOM) of ion chambers in megavoltage beams has been carried out. A one-dimensional scanning phantom system was developed with an overall accuracy in the positioning of a chamber of better than 0.15 mm. Depth-dose data were acquired for a 25 MV beam from an Elekta Precise linac (field sizes of 10x10 cm and 25x25 cm) for measurement depths in the range 0.6-6 cm. The results confirmed the Monte Carlo calculations of an earlier theoretical investigation by Kawrakow [Med. Phys. 33, 1829-1839 (2006)] that the standard shift for cylindrical chambers, recommended in dosimetry protocols of -0.6r (where r is the internal radius of the cavity), is incorrect. A wide range of ion chambers were investigated and it was found that errors of up to 1.4 mm could occur for certain chamber designs (although typical errors for common chambers were around 0.5 mm). A comparison between measurements and Monte Carlo simulations showed that once the correct EPOM is used, the details of the linac geometry are correct, and the parameters of the electron beam striking the bremsstrahlung target have been adequately determined, the EGSnrc Monte Carlo package is capable of reproducing the experimental data to 0.2 mm or better. The investigation also confirmed that for the highest accuracy depth-dose curves in megavoltage photon beams one should use a well-guarded parallel-plate ion chamber. Three chamber designs were tested here and found to be satisfactory--the Scanditronix-Wellhoefer NACP-02, PTW Roos and Exradin All.

OSTI ID:
21036187
Journal Information:
Medical Physics, Vol. 35, Issue 3; Other Information: DOI: 10.1118/1.2839329; (c) 2008 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English