Possibility of using cylindrical ionization chambers for percent depth-dose measurements in clinical electron beams
- Graduate School of Health Sciences, Kumamoto University, 4-24-1 Kuhonji, Kumamoto 862-0976 (Japan)
Purpose: This study investigated the possibility of using cylindrical ionization chambers for percent depth-dose (PDD) measurements in high-energy clinical electron beams. Methods: The cavity correction factor, P{sub cav}, for cylindrical chambers with various diameters was calculated as a function of depth from the surface to R{sub 50}, in the energy range of 6-18 MeV electrons with the EGSnrc C ++ -based user-code CAVITY. The results were compared with those for IBA NACP-02 and PTW Roos parallel-plate ionization chambers. The effective point of measurement (EPOM) for the cylindrical chamber and the parallel-plate chamber was positioned according to the IAEA TRS-398 code of practice. The overall correction factor, P{sub Q}, and the percent depth-ionization (PDI) curve for a PTW30013 Farmer-type chamber were also compared with those of NACP-02 and Roos chambers. Results: The P{sub cav} values at depths between the surface and R{sub 50} for cylindrical chambers were all lower than those with parallel-plate chambers. However, the variation in depth for cylindrical chambers equal to or less than 4 mm in diameter was equivalent to or smaller than that for parallel-plate chambers. The P{sub Q} values for the PTW30013 chamber mainly depended on P{sub cav}, and for parallel-plate chambers depended on the wall correction factor, P{sub wall}, rather than P{sub cav}. P{sub Q} at depths from the surface to R{sub 50} for the PTW30013 chamber was consequently a lower value than that with parallel-plate chambers. However, the variation in depth was equivalent to that of parallel-plate chambers at electron energies equal to or greater than 9 MeV. The shift to match calculated PDI curves for the PTW30013 chamber and water (perturbation free) varied from 0.65 to 0 mm between 6 and 18 MeV beams. Similarly, the shifts for NACP-02 and Roos chambers were 0.5-0.6 mm and 0.2-0.3 mm, respectively, and were nearly independent of electron energy. Conclusions: Calculated PDI curves for PTW30013, NACP-02, and Roos chambers agreed well with that of water by using the optimal EPOM. Therefore, the possibility of using cylindrical ionization chambers can be expected for PDD measurements in clinical electron beams.
- OSTI ID:
- 22100597
- Journal Information:
- Medical Physics, Journal Name: Medical Physics Journal Issue: 8 Vol. 38; ISSN 0094-2405; ISSN MPHYA6
- Country of Publication:
- United States
- Language:
- English
Similar Records
Extraction of depth-dependent perturbation factors for parallel-plate chambers in electron beams using a plastic scintillation detector
Monte Carlo calculations of correction factors for plane-parallel ionization chambers in clinical electron dosimetry
SU-E-T-350: Effective Point of Measurement and Total Perturbation Correction P for Parallel-Plate Ion Chambers in High-Energy Photon Beams
Journal Article
·
Sun Aug 15 00:00:00 EDT 2010
· Medical Physics
·
OSTI ID:22096758
Monte Carlo calculations of correction factors for plane-parallel ionization chambers in clinical electron dosimetry
Journal Article
·
Mon Sep 15 00:00:00 EDT 2008
· Medical Physics
·
OSTI ID:21120911
SU-E-T-350: Effective Point of Measurement and Total Perturbation Correction P for Parallel-Plate Ion Chambers in High-Energy Photon Beams
Journal Article
·
Mon Jun 15 00:00:00 EDT 2015
· Medical Physics
·
OSTI ID:22548401