skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Assessment of Geometrical Accuracy of Multimodal Images Used for Treatment Planning in Stereotactic Radiotherapy and Radiosurgery: CT, MRI and PET

Journal Article · · AIP Conference Proceedings
DOI:https://doi.org/10.1063/1.2356446· OSTI ID:20888200
; ; ; ; ;  [1]
  1. Unidad de Radioneurocirugia, Instituto Nacional de Neurologia y Neurocirugia, Mexico, D.F., 14269 (Mexico)

An acrylic phantom was designed and constructed to assess the geometrical accuracy of CT, MRI and PET images for stereotactic radiotherapy (SRT) and radiosurgery (SRS) applications. The phantom was suited for each image modality with a specific tracer and compared with CT images to measure the radial deviation between the reference marks in the phantom. It was found that for MRI the maximum mean deviation is 1.9 {+-} 0.2 mm compared to 2.4 {+-} 0.3 mm reported for PET. These results will be used for margin outlining in SRS and SRT treatment planning.

OSTI ID:
20888200
Journal Information:
AIP Conference Proceedings, Vol. 854, Issue 1; Conference: 9. Mexican symposium on medical physics, Guadalajara, Jal. (Mexico), 18-21 Mar 2006; Other Information: DOI: 10.1063/1.2356446; (c) 2006 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-243X
Country of Publication:
United States
Language:
English