Title: Quality Assurance of 4D-CT Scan Techniques in Multicenter Phase III Trial of Surgery Versus Stereotactic Radiotherapy (Radiosurgery or Surgery for Operable Early Stage (Stage 1A) Non-Small-Cell Lung Cancer [ROSEL] Study)

Journal Article · · International Journal of Radiation Oncology, Biology and Physics
; ; ;  [1];  [2];  [3];  [1]
  1. Department of Radiation Oncology, VU University Medical Center, Amsterdam (Netherlands)
  2. Department of Radiation Oncology, University Medical Center Groningen, Groningen (Netherlands)
  3. Department of Radiation Oncology, University Medical Center Utrecht, Utrecht (Netherlands)

Purpose: To determine the accuracy of four-dimensional computed tomography (4D-CT) scanning techniques in institutions participating in a Phase III trial of surgery vs. stereotactic radiotherapy (SBRT) for lung cancer. Methods and Materials: All 9 centers performed a 4D-CT scan of a motion phantom (Quasar, Modus Medical Devices) in accordance with their in-house imaging protocol for SBRT. A cylindrical cedar wood insert with plastic spheres of 15 mm (o15) and 30 mm (o30) diameter was moved in a cosine-based pattern, with an extended period in the exhale position to mimic the actual breathing motion. A range of motion of R = 15 and R = 25 mm and breathing period of T = 3 and T = 6 s were used. Positional and volumetric imaging accuracy was analyzed using Pinnacle version 8.1x at various breathing phases, including the mid-ventilation phase and maximal intensity projections of the spheres. Results: Imaging using eight CT scanners (Philips, Siemens, GE) and one positron emission tomography-CT scanner (Institution 3, Siemens) was investigated. The imaging protocols varied widely among the institutions. No strong correlation was found between the specific scan protocol parameters and the observed results. Deviations in the maximal intensity projection volumes averaged 1.9% (starting phase of the breathing cycle [o]15, R = 15), 12.3% (o15, R = 25), and -0.9% (o30, R = 15). The end-expiration volume deviations (13.4%, o15 and 2.5%, o30), were, on average, smaller than the end-inspiration deviations (20.7%, o15 and 4.5%, o30), which, in turn, were smaller than the mid-ventilation deviations (32.6%, o15 and 8.0%, o30). A slightly larger variation in the mid-ventilation origin position was observed (mean, -0.2 mm; range, -3.6-4.2) than in the maximal intensity projection origin position (mean, -0.1 mm; range, -2.5-2.5). The range of motion was generally underestimated (mean, -1.5 mm; range, -5.5-1). Conclusions: Notable differences were seen in the 4D-CT imaging protocols for SBRT among centers. However, the observed deviations in target volumes were generally small. They were slightly larger for the mid-ventilation phases and smallest for the end-expiration phases. Steps to optimize and standardize the 4D-CT scanning protocols for SBRT are desirable.

OSTI ID:
21587555
Journal Information:
International Journal of Radiation Oncology, Biology and Physics, Vol. 80, Issue 3; Other Information: DOI: 10.1016/j.ijrobp.2010.08.017; PII: S0360-3016(10)03063-4; Copyright (c) 2011 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; Country of input: International Atomic Energy Agency (IAEA); ISSN 0360-3016
Country of Publication:
United States
Language:
English