Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

The estimation of the 210Pb bremsstrahlung contribution to the background of lead shielded {gamma}-spectrometers

Journal Article · · AIP Conference Proceedings
DOI:https://doi.org/10.1063/1.2201000· OSTI ID:20798613
; ; ; ; ;  [1]
  1. Department of Physics, Faculty of Sciences, University of Novi Sad, Trg Dositeja Obradovica 4, 21 000 Novi Sad (Serbia and Montenegro)
The 46.5 keV {gamma}-intensity of 210Pb in the background can be easily reduced by inner lining, but the bremsstrahlung from the 1.2 MeV maximal energy {beta}-decay will reach the lead shielded Ge detector. The spectrum of this bremsstrahlung is calculated by numerically fitting the {beta}-spectrum and integrating the Koch-Motz formula. The adsorption of the bremsstrahlung spectrum in the lead is calculated by the effective solid angle algorithm by comparison with the measured background spectrum. It is shown, that for the lead with 25 Bq/kg of 210Pb, at 100 keV the bremsstrahlung contribution to the background is about 20%.
OSTI ID:
20798613
Journal Information:
AIP Conference Proceedings, Journal Name: AIP Conference Proceedings Journal Issue: 1 Vol. 831; ISSN APCPCS; ISSN 0094-243X
Country of Publication:
United States
Language:
English

Similar Records

The 210Pb Bremsstrahlung In The Background Of Lead Shields
Journal Article · Mon Apr 23 00:00:00 EDT 2007 · AIP Conference Proceedings · OSTI ID:21057297

Recent Bremsstrahlung-based Assays of 210Pb in Lead and Comments on Current Availability of Low-Background Lead in North America
Journal Article · Tue Aug 01 00:00:00 EDT 2017 · Applied Radiation and Isotopes · OSTI ID:1608103

Assay Methods for 238U, 232Th, and 210Pb in Lead and Calibration of 210Bi Bremsstrahlung Emission from Lead