A 100 % Duty-ratio 3 x 6 Arrayed CCD X-ray Detector for Rapid Data Collection in Macromolecular Crystallography

Journal Article · · AIP Conference Proceedings
DOI:https://doi.org/10.1063/1.1757944· OSTI ID:20653103
 [1];  [2];  [3];  [4]
  1. Institute of Materials Science, University of Tsukuba, 1-1-1 Tennodai, Tsukuba, 305-8573 Ibaraki (Japan)
  2. Photon Factory, Institute of Materials Structure Science, KEK, 1-1 Oho, Tsukuba, 305-0801 Ibaraki (Japan)
  3. Foundation for Advancement of International Science (FAIS), PF/IMSS KEK, Tsukuba, 305-0801 Ibaraki (Japan)
  4. Graduate School of Frontier Sciences, University of Tokyo, 5-1-5 Kashiwanoha, Kashiwa, 277-8561 Chiba (Japan)

A 3 x 6 arrayed charge coupled device (CCD) X-ray detector has been developed for rapid data collection in macromolecular crystallography at Photon Factory. The total active area is 236 mm x 236 mm (76.8 {mu}m x 76.8 {mu}m / pixel at the input phosphor, 3072 x 3072 pixels). Each module consists of an input phosphor, a 3.2 : 1 fiber- optic-taper (FOT), a CCD, and a readout circuit. The input phosphor is Gd2O2S:Tb (P-43), which is deposited onto an aluminum coated Mylar sheet to increase the sensitivity. The outputs from eighteen (3 x 6) CCDs are readout in parallel and then are digitized by 16 bit analog-to-digital converters. The advantage of this detector over the conventional FOT-coupled CCD detectors is a unique CCD readout scheme that enables us to record successive x-ray exposures without interruption among them. A shutter is not needed for CCD readout and the sample crystal can be rotated continuously without interruption during data acquisition. We have collected one data set of a Lysozyme crystal for 180 degree with 1.5 degree per slice in 360 seconds. The exposure time of each slice is 3 seconds. The duty-ratio of the x-ray exposure to the data collection time is 100 %. This detector combined with a synchrotron radiation source is very suitable for rapid data collection in macromolecular crystallography.

OSTI ID:
20653103
Journal Information:
AIP Conference Proceedings, Journal Name: AIP Conference Proceedings Journal Issue: 1 Vol. 705; ISSN APCPCS; ISSN 0094-243X
Country of Publication:
United States
Language:
English

Similar Records

Fiber-Optic Taper Coupled with a Large Format Charge-coupled Device X-ray Detector: Fast Readout and High Duty-Cycle Ratio
Journal Article · Thu Jan 18 23:00:00 EST 2007 · AIP Conference Proceedings · OSTI ID:21049228

X-ray area detector based on CCD readout for diffraction study (abstract)
Journal Article · Tue Dec 31 23:00:00 EST 1991 · Review of Scientific Instruments; (United States) · OSTI ID:5520103

Optimal fine ϕ-slicing for single-photon-counting pixel detectors