Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Integrating amplifiers for PHENIX lead-glass and lead-scintillator calorimeters

Conference ·
OSTI ID:201764
; ;  [1]; ;  [2]
  1. Oak Ridge National Lab., TN (United States)
  2. Univ. of Tennessee, Knoxville, TN (United States)
Two types of integrating amplifier systems have been developed for use with lead-glass and lead-scintillator calorimeters with photomultiplier tube readout. Requirements for the amplifier system include termination of the line from the photomultiplier, compact size and low power dissipation to allow multiple channels per chip, dual range outputs producing 10-bit accuracy over a 14-bit dynamic range, rms noise levels of one LSB or less, and compatibility with timing filter amplifiers, tower sum circuits for triggering and calibration circuits to be built on the same integrated circuit (IC). Advantages and disadvantages of an active integrator system are compared and contrasted to those of a passive integrator-based system. In addition, details of the designs and results from prototype devices including an 8-channel active integrator IC fabricated in 1.2 {micro}m Orbit CMOS are presented.
Research Organization:
Oak Ridge National Lab., TN (United States)
Sponsoring Organization:
USDOE, Washington, DC (United States)
DOE Contract Number:
AC05-84OR21400
OSTI ID:
201764
Report Number(s):
CONF-951073--20; ON: DE96005484
Country of Publication:
United States
Language: